FREE PSA

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2015-02-26 for FREE PSA manufactured by Roche Diagnostics.

Event Text Entries

[18034443] The customer received questionable total psa total (free + complexed) psa - prostate-specific antigen (total psa) and free psa results for an unknown number of patient samples. Of the data provided, only the data for one patient was discrepant and did not agree with the clinical status of the patient. (b)(4). The initial results for the patient from cobas e411 serial number (b)(4) were total psa: 94. 42 ng/ml and free psa: 15. 25 ng/ml. The urologist questioned these results and sent the sample to be tested at three other sites. Site 1 (modular e170) on (b)(6) 2015: total psa: 90. 19 ng/ml free psa: 22. 69 ng/ml. Site 2 (cobas e411) on (b)(6) 2015: total psa: 73. 99 ng/ml free psa: 13 ng/ml. Site 3 (cobas e411) on (b)(6) 2015: total psa: 67. 67 ng/ml free psa: 13. 77 ng/ml. The patient was not adversely affected.
Patient Sequence No: 1, Text Type: D, B5


[18134187] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[18442893] A specific root cause could not be identified as the sample was not available for further testing. Additional information for further investigation such as which result was believed to be correct was requested but was not provided. Review of the provided calibration and qc data did not indicate a general reagent or system issue.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2015-01576
MDR Report Key4549117
Report Source01,05,06
Date Received2015-02-26
Date of Report2015-04-15
Date of Event2015-02-06
Date Mfgr Received2015-02-10
Date Added to Maude2015-02-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameFREE PSA
Generic NameTEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA
Product CodeMTG
Date Received2015-02-26
Model NumberNA
Catalog NumberASKU
Lot NumberASKU
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2015-02-26

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.