XOMED? INSTRUMENTS 3722071

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07 report with the FDA on 2015-03-06 for XOMED? INSTRUMENTS 3722071 manufactured by Medtronic Xomed, Inc..

Event Text Entries

[5557173] It was reported that during a procedure, the fine tip of the product broke off into the patient. She stated the fragment was picked up? Pretty easily? And removed from the patient. There was no injury reported as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


[13024727] (b)(6). (b)(4). The product analysis indicates that the tip was broken off. The fragment was also returned. There was no evidence of any material defects that would have resulted in the break. The most likely cause of the reported event is misuse/user error. This device is used for therapeutic purposes.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1045254-2015-00066
MDR Report Key4576287
Report Source05,06,07
Date Received2015-03-06
Date of Report2015-02-11
Date of Event2015-01-11
Date Mfgr Received2015-02-11
Date Added to Maude2015-04-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactURIZA SHUMS
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal32216
Manufacturer Phone9043328405
Manufacturer G1MEDTRONIC XOMED, INC.
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal Code32216
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameXOMED? INSTRUMENTS
Generic NamePICK, MICROSURGICAL EAR
Product CodeJYT
Date Received2015-03-06
Returned To Mfg2015-02-17
Model Number3722071
Catalog Number3722071
Lot NumberNI
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC XOMED, INC.
Manufacturer Address6743 SOUTHPOINT DR N JACKSONVILLE FL 32216 US 32216


Patients

Patient NumberTreatmentOutcomeDate
10 2015-03-06