HOYER STATURE HOY-STATURE

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2015-02-26 for HOYER STATURE HOY-STATURE manufactured by Apex Healthcare Mfg Inc.

Event Text Entries

[5561157] It was reported to the manufacturer by the facility (b)(6), per the facility 2 cna's were transferring a resident with the hoyer stature when it tipped over. The resident landed on one of the cna's and caused cna head to hit bedside table and floor. The other cna sustained injuries to her leg when she also fell. Both cna's were sent to the urgent care clinic. The cna's have hematoma on head and leg as well as swelling to the head, shoulder and leg. A joerns representative visited the facility on (b)(6) 2015. The lift and sling were evaluated. The casters on the lift were "sticking", making it more difficult to move the lift. The facility ordered new casters for all their lifts. During the incident, training provided by joerns rep. It was admitted that the cna put her foot on the leg of the lift and pulled the resident back into the chair by pulling on the sling. This caused the lift to tip over onto the cna's. It was also mentioned that the staff had tipped the lift in the past, but it did not go all the way over and was not reported to management until this incident when the lift went all the way over. Complaint # (b)(4) was entered into our system.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3009402404-2015-00005
MDR Report Key4577973
Report Source06
Date Received2015-02-26
Date of Report2015-02-27
Date of Event2015-02-09
Date Facility Aware2015-02-09
Report Date2015-02-24
Date Reported to FDA2015-02-24
Date Reported to Mfgr2015-02-24
Date Mfgr Received2015-02-09
Device Manufacturer Date2014-01-01
Date Added to Maude2015-03-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactFELICIA BANKS
Manufacturer Street2100 DESIGN RD
Manufacturer CityARLINGTON TX 76014
Manufacturer CountryUS
Manufacturer Postal76014
Manufacturer Phone8008260270
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameHOYER STATURE
Generic NameSIT TO STAND LIFT
Product CodeFNG
Date Received2015-02-26
Model NumberHOY-STATURE
Catalog NumberHOY-STATURE
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Age1 YR
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerAPEX HEALTHCARE MFG INC
Manufacturer AddressMIN HSIUNG CHI YI TW


Patients

Patient NumberTreatmentOutcomeDate
10 2015-02-26

© 2026 FDA.report