75FT (5-PIN) DC EXTENSION CORD 0240030952

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2015-03-06 for 75FT (5-PIN) DC EXTENSION CORD 0240030952 manufactured by Stryker Endoscopy-san Jose.

Event Text Entries

[5558539] It was reported the while testing a power extension a blue spark flashed from the while 4 pin short adaptor cable block. The monitor failed to power up.
Patient Sequence No: 1, Text Type: D, B5


[12955684] Additional information will be provided once the investigation has been completed.
Patient Sequence No: 1, Text Type: N, H10


[31145916] The product was returned and the failure mode was confirmed. Visual inspection: there is minor scratches to the 75ft (5-pin) dc extension cord. The soldering of the wires does not match when the leads are connected. Functional inspection : the soldering of the wires does not match when the leads are connected. No power was duplicated during test with a known working monitor. The issue was duplicated. Probable root cause is assembly. The product was returned for investigation and the reported failure mode was confirmed. The failure mode will be monitored for future reoccurrence.
Patient Sequence No: 1, Text Type: N, H10


[31145917] It was reported the while testing a power extension a blue spark flashed from the while 4 pin short adaptor cable block. The monitor failed to power up.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number0002936485-2015-00125
MDR Report Key4579223
Report Source07
Date Received2015-03-06
Date of Report2015-02-11
Date of Event2015-02-11
Date Mfgr Received2015-02-11
Date Added to Maude2015-04-07
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. THOMAS SHAFER
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal95138
Manufacturer Phone4087542000
Manufacturer G1STRYKER ENDOSCOPY-SAN JOSE
Manufacturer Street5900 OPTICAL COURT
Manufacturer CitySAN JOSE CA 95138
Manufacturer CountryUS
Manufacturer Postal Code95138
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name75FT (5-PIN) DC EXTENSION CORD
Generic NameCAMERA, TELEVISION, SURGICAL, WITHOUT AUDIO
Product CodeFWB
Date Received2015-03-06
Returned To Mfg2015-03-12
Catalog Number0240030952
Lot Number201402
OperatorOTHER
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER ENDOSCOPY-SAN JOSE
Manufacturer Address5900 OPTICAL COURT SAN JOSE CA 95138 US 95138


Patients

Patient NumberTreatmentOutcomeDate
10 2015-03-06

© 2026 FDA.report