MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,01,05 report with the FDA on 2015-03-09 for VITROS CHEMISTRY PRODUCTS DGXN SLIDES 8343386 manufactured by Ortho-clinical Diagnostics.
[5480922]
A customer observed lower than expected dgxn quality control results from omnicore qc fluids when using vitros chemistry products dgxn slides on a vitros 5600 integrated system. Omnicore l2: dgxn result 0. 55 ng/ml vs. Expected result 1. 68 ng/ml. Omnicore l3: dgxn results 0. 58, 0. 47, <0. 4, <0. 4, 0. 51 ng/ml vs. Expected result 2. 00 ng/ml. Biased results of the magnitude and direction observed may lead to inappropriate physician action. No patient samples were processed on this analyzer; however, the investigation cannot conclude that patient samples would not be affected if the event were to recur undetected. There was no allegation of patient harm as a result of this event. This report is number six of six mdr? S for this event. Six 3500a forms are being submitted for this event as six devices were involved. This report corresponds to ortho clinical diagnostics inc. (b)(4).
Patient Sequence No: 1, Text Type: D, B5
[13026377]
The investigation confirmed that lower than expected dgxn quality control results were obtained from omnicore qc fluids when using vitros chemistry products dgxn slides on a vitros 5600 integrated system. The investigation was unable to determine a definitive assignable cause. The omnicore quality control fluids could not be ruled out as a contributing factor. The investigation found no evidence to suggest unexpected vitros dgxn slide or vitros 5600 instrument performance contributed to the event.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 1319809-2015-00011 |
MDR Report Key | 4583165 |
Report Source | 00,01,05 |
Date Received | 2015-03-09 |
Date of Report | 2015-03-09 |
Date of Event | 2015-02-11 |
Date Mfgr Received | 2015-02-11 |
Device Manufacturer Date | 2014-11-02 |
Date Added to Maude | 2015-04-08 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | MR. JAMES A STEVENS |
Manufacturer Street | 100 INDIGO CREEK DRIVE |
Manufacturer City | ROCHESTER NY 14626 |
Manufacturer Country | US |
Manufacturer Postal | 14626 |
Manufacturer Phone | 5854533000 |
Manufacturer G1 | ORTHO-CLINICAL DIAGNOSTICS |
Manufacturer Street | 513 TECHNOLOGY BLVD. |
Manufacturer City | ROCHESTER NY 14652 |
Manufacturer Country | US |
Manufacturer Postal Code | 14652 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | VITROS CHEMISTRY PRODUCTS DGXN SLIDES |
Generic Name | IN-VITRO DIAGNOSTIC |
Product Code | LFM |
Date Received | 2015-03-09 |
Catalog Number | 8343386 |
Lot Number | 1906-0227-8610 |
Device Expiration Date | 2015-04-01 |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ORTHO-CLINICAL DIAGNOSTICS |
Manufacturer Address | 100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2015-03-09 |