R. WOLF CO2 YOKE PART# 2026-64 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2003-05-05 for R. WOLF CO2 YOKE PART# 2026-64 * manufactured by Richard Wolf Medical Instruments.

Event Text Entries

[22072666] The short indexing pins allowed a nitrous oxide bottle to be clamped in a co2 yoke for insufflation without a leak. Therefore, nitrous oxide was insufflated into the pt's abdomen.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW1028350
MDR Report Key458651
Date Received2003-05-05
Date of Report2003-05-05
Date of Event2003-04-01
Date Added to Maude2003-05-12
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA0
Report to FDA0
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameR. WOLF CO2 YOKE
Generic NameYOKE/CLAMP
Product CodeCAM
Date Received2003-05-05
Model NumberPART# 2026-64
Catalog Number*
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key447614
ManufacturerRICHARD WOLF MEDICAL INSTRUMENTS
Manufacturer Address353 CORPORATE WOODS PARKWAY VERNON HILLS IL 600613110 US


Patients

Patient NumberTreatmentOutcomeDate
10 2003-05-05

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.