VITROS CHEMISTRY PRODUCTS CRBM SLIDES 8892382

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2015-03-11 for VITROS CHEMISTRY PRODUCTS CRBM SLIDES 8892382 manufactured by Ortho-clinical Diagnostics.

Event Text Entries

[5342759] The customer complained that higher than expected vitros crbm results were obtained from a single level of vitros tdm fluid tested on a 5600 integrated system. Tdm pv l3673: 17. 53, 17. 76, 17. 60 and 17. 49 ug/ml vs expected 13. 32 ug/ml. Biased results of the magnitude and direction observed may lead to inappropriate physician action. The customer made no indication that patient samples were affected, however, the investigation could not rule out that patient samples were not affected, or would not be affected if the event were to recur undetected. There was no allegation of patient harm as a result of this event. This report is number two of four mdr? S for this event. Four 3500a forms are being submitted for this event as four devices were involved. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[13013108] The investigation determined that higher than expected vitros crbm results were obtained from a single level of vitros tdm fluid tested on a 5600 integrated system. The assignable cause of the event is unknown; however, an unexpected instrument issue or user error due to inappropriate pre-analytical sample handling cannot be ruled out as contributing factors. There was no indication the vitros crbm micro slide malfunctioned.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1319809-2015-00029
MDR Report Key4590577
Report Source05
Date Received2015-03-11
Date of Report2015-03-11
Date of Event2015-02-16
Date Mfgr Received2015-02-16
Device Manufacturer Date2014-11-13
Date Added to Maude2015-04-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. JAMES A STEVENS
Manufacturer Street100 INDIGO CREEK DRIVE
Manufacturer CityROCHESTER NY 14626
Manufacturer CountryUS
Manufacturer Postal14626
Manufacturer Phone5854533000
Manufacturer G1ORTHO-CLINICAL DIAGNOSTICS
Manufacturer Street513 TECHNOLOGY BLVD.
Manufacturer CityROCHESTER NY 14652
Manufacturer CountryUS
Manufacturer Postal Code14652
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVITROS CHEMISTRY PRODUCTS CRBM SLIDES
Generic NameIN-VITRO DIAGNOSTIC
Product CodeKLT
Date Received2015-03-11
Catalog Number8892382
Lot Number3924-0085-8851
Device Expiration Date2015-07-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerORTHO-CLINICAL DIAGNOSTICS
Manufacturer Address100 INDIGO CREEK DRIVE ROCHESTER NY 14626 US 14626


Patients

Patient NumberTreatmentOutcomeDate
10 2015-03-11

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.