AIRBORNE 750I INFANT *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-03-06 for AIRBORNE 750I INFANT * manufactured by International Biomedical, Ltd..

Event Text Entries

[20364282] A newborn was in transit to the nicu in the device. As the nurse entered the room with the device, the top portion of the device started to slide off the base. The nurse immediately stopped and removed the newborn from the device. The newborn was transported in arms the rest of the way into the room and placed in the warmer. The device was removed from service and sent to biomed for examination. Biomed determined that the 4 bolts that hold the top portion of the device had come loose. The loose bolts were replaced. The bolts are located near shock absorbers, which may have contributed to the loose bolts. Our biomed team determined that the bolts should be added to the 6 month preventative maintenance schedule for this device.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number4591765
MDR Report Key4591765
Date Received2015-03-06
Date of Report2015-03-06
Date of Event2014-12-07
Report Date2015-03-06
Date Reported to FDA2015-03-06
Date Reported to Mfgr2015-03-11
Date Added to Maude2015-03-11
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameAIRBORNE 750I INFANT
Generic NameINCUBATOR, NEONATAL TRANSPORT
Product CodeFPL
Date Received2015-03-06
Model Number750I
Catalog Number*
Lot Number*
ID Number*
OperatorNURSE
Device AvailabilityN
Device Sequence No1
Device Event Key0
ManufacturerINTERNATIONAL BIOMEDICAL, LTD.
Manufacturer Address8508 CROSS PARK DRIVE AUSTIN TX 78754 US 78754


Patients

Patient NumberTreatmentOutcomeDate
10 2015-03-06