PRIMUS FLEXIBLE GREAT TOE IMPLANT, SIZE 30 FGT-30

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 99,company representative,unk report with the FDA on 2015-03-05 for PRIMUS FLEXIBLE GREAT TOE IMPLANT, SIZE 30 FGT-30 manufactured by Tornier Inc..

Event Text Entries

[5628418] The implant appears to have subsided/shifted "plantarly" in the toe.
Patient Sequence No: 1, Text Type: D, B5


[13049185] This is the initial report submitted regarding this surgical event and medical device.
Patient Sequence No: 1, Text Type: N, H10


[32716607] Repeated attempts made to surgeon and complaint reporter to obtain x-rays and additional information about the case were unsuccessful. There appears to be no defect in the device or manufacturing of the product returned. Due to an inability to obtain an x-ray or additional patient information it is not possible to determine if the original implant was seated properly in the bone. Additionally, er were unable to obtain any additional information regarding bone density or other possible contributing factors from the surgeon. Without additional information the root cause of this issue cannot be determined. This is the final report submitted regarding this surgical event and medical device.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004983210-2015-00003
MDR Report Key4597357
Report Source99,COMPANY REPRESENTATIVE,UNK
Date Received2015-03-05
Date of Report2015-02-11
Date Mfgr Received2015-02-11
Date Added to Maude2015-04-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactKEVIN SMITH
Manufacturer Street10801 NESBITT AVE., SOUTH
Manufacturer CityBLOOMINGTON MN 55437
Manufacturer CountryUS
Manufacturer Postal55437
Manufacturer Phone9524267643
Manufacturer G1TORNIER INC.
Manufacturer Street10801 NESBITT AVE., SOUTH
Manufacturer CityBLOOMINGTON MN 55437
Manufacturer CountryUS
Manufacturer Postal Code55437
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePRIMUS FLEXIBLE GREAT TOE IMPLANT, SIZE 30
Generic NameNONE
Product CodeKWH
Date Received2015-03-05
Catalog NumberFGT-30
Lot NumberCT0414140
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTORNIER INC.
Manufacturer AddressBLOOMINGTON MN US


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2015-03-05

© 2026 FDA.report