PROVOX VEGA 8131

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2015-03-13 for PROVOX VEGA 8131 manufactured by Atos Medical Ab.

Event Text Entries

[5628441] Suddenly speaking problems occurred. Patient checked vp in mirror and noticed that it was broken. Since the visible part of the vp was broken/torn off and almost fell into the trachea, he himself drew out the vp immediately.
Patient Sequence No: 1, Text Type: D, B5


[12977803] This report is a correction of 8032044-2015-00004 which by mistake was classified as a 5 day report. Investigation: the vega has an xtraflange mounted. It is actually the esophagus flange that is ruptured (not visible to the patient). The tracheal flange is undamaged. The complaint description does not match the damage on the voice prosthesis. The esophagus flange is ruptured about 80% of the circumference, about 20% is still attached, no parts are missing. The fraction surface does not have any enclosed air bubbles that could have caused the tracheal flange to rupture. There are some nicks and scratches indicating that the outside of the housing has been in contact with a sharp tool of some kind. A possible scenario is that wrong method and/or tool was used while mounting the xtraflange and causing the scratch on the housing and at the same time an indication of fraction on the esophagus flange as well. This fraction then resulted in a ruptured esophagus flange at the time of removal of the voice prosthesis. The blue ring is still safely attached to the waist of the prosthesis. Conclusion: judged from the complaint description, damage on the housing and the surface of the rupture it is not likely that this rupture is a result of a faulty product.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8032044-2015-00005
MDR Report Key4598202
Report Source07
Date Received2015-03-13
Date of Report2015-02-09
Date Mfgr Received2014-12-19
Device Manufacturer Date2014-09-01
Date Added to Maude2015-04-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR FERENC DAHNER
Manufacturer StreetPO-BOX 183
Manufacturer CityHORBY, SE-242 22
Manufacturer CountrySW
Manufacturer PostalSE-242 22
Manufacturer Phone415 19800
Manufacturer G1ATOS MEDICAL AB
Manufacturer StreetPO BOX 183 KRAFTGATAN 8
Manufacturer CityHORBY, SE-242 2
Manufacturer CountrySW
Manufacturer Postal CodeSE-242 22
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePROVOX VEGA
Generic NameVOICE PROSTHESIS
Product CodeEWL
Date Received2015-03-13
Returned To Mfg2015-02-02
Catalog Number8131
Lot Number1409045
Device Expiration Date2019-08-30
OperatorLAY USER/PATIENT
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerATOS MEDICAL AB
Manufacturer AddressPO BOX 183 KRAFTGATAN 8 HORBY, SE-242 2 SW SE-242 22


Patients

Patient NumberTreatmentOutcomeDate
10 2015-03-13

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.