RISK MANAGER 37052

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,06 report with the FDA on 2015-03-13 for RISK MANAGER 37052 manufactured by Span America Medical Systems, Inc..

Event Text Entries

[5467771] An employee was walking into a patients room and stepped one foot onto the risk manager (a bedside safety mat). As she stepped on the mat, it slipped, causing her legs to split apart. The slip of the mat caused a groin injury to the employee and she is now on light duty.
Patient Sequence No: 1, Text Type: D, B5


[13002924] The risk manager is a fall pad designed to reduce the likelihood of a serious injury in the event of a fall at patient bad side. This device was not returned since there was not an obvious problem with the product. This product can become slick depending on the floor surface, floor cleaner and application (mat placed on wet surface), etc. We do sell "grip strips" which are available for anyone that finds a need which is included in our product insert. The facility was not using the grip strips so we have sent 5 sets for their evaluation.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1041130-2015-00001
MDR Report Key4599678
Report Source01,06
Date Received2015-03-13
Date of Report2015-03-11
Date of Event2015-02-22
Date Mfgr Received2015-03-02
Date Added to Maude2015-03-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactWILLIAM DARBY
Manufacturer Street70 COMMERCE CENTER
Manufacturer CityGREENVILLE SC 29615
Manufacturer CountryUS
Manufacturer Postal29615
Manufacturer Phone8646786936
Manufacturer G1SPAN AMERICA MEDICAL SYSTEMS, INC.
Manufacturer Street70 COMMERCE CENTER
Manufacturer CityGREENVILLE SC 29615
Manufacturer CountryUS
Manufacturer Postal Code29615
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRISK MANAGER
Generic NameBEDSIDE SAFETY MAT - FALL MAT
Product CodeFMP
Date Received2015-03-13
Model Number37052
Catalog Number37052
OperatorDEVICE UNATTENDED
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSPAN AMERICA MEDICAL SYSTEMS, INC.
Manufacturer Address70 COMMERCE CENTER GREENVILLE SC 29615 US 29615


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-03-13