FREE PSA 03289788190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2015-03-13 for FREE PSA 03289788190 manufactured by Roche Diagnostics.

Event Text Entries

[5696115] The customer stated that they obtained erroneous results for two samples from one patient tested for total (free + complexed) psa - prostate-specific antigen (tpsa) and free psa (fpsa) on an e601 analyzer. The fpsa results of the patient were greater than the tpsa results. The erroneous results from one of the two patient samples were reported outside of the laboratory. This medwatch will cover the fpsa assay. (b)(4). The patient sample resulted as 2. 62 ng/ml for fpsa and 0. 62 ng/ml for tpsa. The patient was not adversely affected. The e601 analyzer serial number was (b)(4). A specific root cause could not be determined based on the information provided. A sample from the patient was requested for investigation purposes, but could not be provided.
Patient Sequence No: 1, Text Type: D, B5


[13010203] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2015-02027
MDR Report Key4599978
Report Source01,05,06
Date Received2015-03-13
Date of Report2015-03-13
Date of Event2014-08-19
Date Mfgr Received2015-02-19
Date Added to Maude2015-03-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameFREE PSA
Generic NameTEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA
Product CodeMTG
Date Received2015-03-13
Model NumberNA
Catalog Number03289788190
Lot NumberASKU
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2015-03-13