TRANSMOTION MEDICAL, INC. TMM5-TB

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2015-03-10 for TRANSMOTION MEDICAL, INC. TMM5-TB manufactured by Transmotion Medical Inc..

Event Text Entries

[5629000] Customer called technical service line and said that the chair moved on its own without pushing any of the buttons on the pendant. Email received from (b)(6) said the chair moved during a procedure but there were no injuries caused at all.
Patient Sequence No: 1, Text Type: D, B5


[12975673] The immediate action taken was to replace older style pendants with newly designed pendants for all devices in the facility under the current voluntary recall. There was not an adverse event. The customer originally said that the chair moved on its own. Then later in an email, said the chair moved during a procedure "no injuries caused at all. " transmotion medical inc has designed a new pendant which was released in may of 2013. The new pendant is considerably more durable than the older style pendants. The new pendant has been in production for 22 months and there has been one customer claim that the new pendant caused the chair to move on its own but this could not be confirmed. In november 2014, transmotion medical, inc launched a product recall due to previous customer complaint and corresponding mdr data. This mdr is for voluntary reporting purposes only.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004082462-2015-00001
MDR Report Key4600679
Report Source05
Date Received2015-03-10
Date of Report2015-03-10
Date of Event2015-02-12
Date Mfgr Received2015-02-12
Date Added to Maude2015-03-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street1441 WOLF CREEK TRAIL
Manufacturer CitySHARON CENTER OH 44274
Manufacturer CountryUS
Manufacturer Postal44274
Manufacturer Phone3302394192
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTRANSMOTION MEDICAL, INC.
Generic NameTMM5 SURGICAL STRETCHER-CHAIR
Product CodeGBB
Date Received2015-03-10
Model NumberTMM5-TB
Catalog NumberTMM5-TB
Lot Number0435
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerTRANSMOTION MEDICAL INC.
Manufacturer Address1441 WOLF CREEK TRAIL SHARON CENTER OH 44274 US 44274


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-03-10