MS CEILING LIFT, MAXISKY600 MS600

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-03-19 for MS CEILING LIFT, MAXISKY600 MS600 manufactured by Arjohuntleigh.

Event Text Entries

[5767525] During annual preventive maintenance inspection on ms600 biomed techs found that the emergency stop failed to work. When the ms600 was lowering 591 lbs and the emergency pull cord was engaged (pulled) the lift would accelerate down and continue to pick up speed. When the cord was pulled again the lift would slow down but would continue to move towards the ground. The lift was then tested with 200 lbs and when the emergency cable was pulled the lift would shut down and come to a stop. The lift is rated for 600 lbs and should not change function for any weight user that amount. This was tested on other lifts at our facility. The new lifts which have the new style emergency stop functioned properly. The problem listed above was only found on the old style lifts with pull string that are similar to a ceiling fan pull string. If there are any questions or any need for further info please contact me directly at the number listed below.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5041576
MDR Report Key4627506
Date Received2015-03-19
Date of Report2015-02-17
Date of Event2015-02-11
Date Added to Maude2015-03-25
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameMS CEILING LIFT, MAXISKY600
Generic NameMS CEILING LIFT, MAXISKY600
Product CodeFNG
Date Received2015-03-19
Model NumberMS600
OperatorOTHER
Device AvailabilityY
Device Sequence No1
Device Event Key0
ManufacturerARJOHUNTLEIGH
Manufacturer Address2349, W LAKE STREET, SUITE 250 ADDISON IL 60101 US 60101


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2015-03-19

© 2026 FDA.report