MONOMAX VIOLET 0 (3,5) 150CM HR26 (M) B0041441

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,07 report with the FDA on 2015-02-27 for MONOMAX VIOLET 0 (3,5) 150CM HR26 (M) B0041441 manufactured by B Braun Surgical S.a..

Event Text Entries

[5601066] Country of complaint: (b)(6). Inner foil was welded with outer foil. Due to this fact a sterile withdrawal was not possible.
Patient Sequence No: 1, Text Type: D, B5


[13043877] Us reporting agent notified on: (b)(6)2015. Manufacturing site evaluation: samples received: 26 unopened pouches and 1 opened. There are no previous complaints of this code batch. There are no units in oem stock. Received 26 closed samples and one opened (only the second pack is opened and with the aluminum pack stuck to the paper foil). This defect is caused in the welding machine during the manufacturing process of the product. Final conclusion: the complaint is corresponding (justified). Corrective/preventive actions: not applicable.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2916714-2015-00149
MDR Report Key4645900
Report Source01,07
Date Received2015-02-27
Date of Report2015-02-27
Date Mfgr Received2015-01-12
Device Manufacturer Date2014-11-01
Date Added to Maude2015-05-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactNICOLE BROYLES
Manufacturer Street615 LAMBERT POINTE DRIVE
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3145515988
Manufacturer G1B BRAUN SURGICAL SA
Manufacturer Street121 CARRETERA DE TERRASSA
Manufacturer CityRUBI BARCELONA
Manufacturer CountrySP
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMONOMAX VIOLET 0 (3,5) 150CM HR26 (M)
Generic NameSYNTHETIC ABSORBABLE SUTURE
Product CodeNWJ
Date Received2015-02-27
Model NumberB0041441
Catalog NumberB0041441
Lot Number11/01/2019
OperatorHEALTH PROFESSIONAL
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerB BRAUN SURGICAL S.A.
Manufacturer AddressRUBI (BARCELONA) SP


Patients

Patient NumberTreatmentOutcomeDate
10 2015-02-27

© 2026 FDA.report