UNICEL DXC 600 PRO SYNCHRON SYSTEM A11810

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2015-03-31 for UNICEL DXC 600 PRO SYNCHRON SYSTEM A11810 manufactured by Beckman Coulter.

Event Text Entries

[5599641] Customer reported the unicel dxc 600 pro synchron system generated erratic and suppressed results for blood urea nitrogen (bun) for patient and quality control samples. Customer also reported reagent probe a was leaking. Customer reported the leak was about 15ml and contained to under the home position of the reagent probe a. No erroneous bun results were reported out of the lab and no change in treatment or harm to patients. No exposure reported. Customer was wearing gloves.
Patient Sequence No: 1, Text Type: D, B5


[13099053] A field service engineer (fse) was dispatched to evaluate the instrument. Fse determined the cause of the erratic and suppressed bun patient and qc results as well as the reagent probe a leak was from reagent probe a collar wash valve. Service replaced the collar wash valve and that resolved the issue.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2015-00115
MDR Report Key4646552
Report Source05,06
Date Received2015-03-31
Date of Report2015-03-07
Date of Event2015-03-07
Date Mfgr Received2015-03-07
Device Manufacturer Date2007-06-25
Date Added to Maude2015-05-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. DAVID DAVIS
Manufacturer Street250 S. KRAEMER BLVD. M/S E1.SE.01
Manufacturer CityBREA CA 928218000
Manufacturer CountryUS
Manufacturer Postal928218000
Manufacturer Phone7142649714
Manufacturer G1BECKMAN COULTER
Manufacturer Street250 S. KRAEMER BLVD.
Manufacturer CityBREA CA 92821800
Manufacturer CountryUS
Manufacturer Postal Code92821 8000
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameUNICEL DXC 600 PRO SYNCHRON SYSTEM
Generic NameANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE
Product CodeCDN
Date Received2015-03-31
Model NumberNA
Catalog NumberA11810
Lot NumberNA
ID NumberSOFTWARE VERSION 5.3
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER
Manufacturer Address250 S. KRAEMER BLVD. BREA CA 92821800 US 92821 8000


Patients

Patient NumberTreatmentOutcomeDate
10 2015-03-31

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.