ELITE CARE CLINER W/SWING AWAY ARMS 6940

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2015-03-30 for ELITE CARE CLINER W/SWING AWAY ARMS 6940 manufactured by Winco Mfg Llc.

Event Text Entries

[16850221] On (b)(6) 2014 winco customer svc rec'd a call from (b)(6), area tech operations mgr for fresenius medical care. He informed us that a 6940 elite care cliner had overheated during a pt treatment and left a burn pattern on the chair back cushion and the pts pillow. The pt initially had red marks on his back similar to a sunburn. The pt refused to go to the hosp. Upon completion of dialysis, the pt had only a small red mark on his back. Pt rec'd no medical care for the red marks (b)(4). The facility refused repeated requests from winco to return the recliner for eval for approximately one month.
Patient Sequence No: 1, Text Type: D, B5


[17081111] The facility contacted winco about the incident, by phone, on 12/31/2014 but refused to return the chair to winco for eval until 01/29/2015. This is the "becomes aware" date that we used. On 02/27/2015, advised clinic to inspect their remaining recliners for worn or incorrectly installed wiring harness.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1027229-2015-00001
MDR Report Key4647435
Report Source07
Date Received2015-03-30
Date of Report2015-02-27
Date of Event2014-12-31
Date Mfgr Received2014-12-31
Device Manufacturer Date2010-11-19
Date Added to Maude2015-05-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJAIME ACEVEDO
Manufacturer Street5516 S.W. FIRST LN.
Manufacturer CityOCALA FL 34474
Manufacturer CountryUS
Manufacturer Postal34474
Manufacturer Phone3528542926
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameELITE CARE CLINER W/SWING AWAY ARMS
Generic NameCLINICAL RECLINER, 6940-19-TB-HM
Product CodeFRJ
Date Received2015-03-30
Model Number6940
Lot NumberNA
ID NumberNA
OperatorLAY USER/PATIENT
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerWINCO MFG LLC
Manufacturer Address5516 S.W. FIRST LN. OCALA FL 34474 US 34474


Patients

Patient NumberTreatmentOutcomeDate
10 2015-03-30