RUSCH MRI COND FIBER OPTIC LARYN KIT 005852300

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06,07 report with the FDA on 2015-03-27 for RUSCH MRI COND FIBER OPTIC LARYN KIT 005852300 manufactured by Teleflex Medical.

Event Text Entries

[5599821] The customer alleges that the mac 2 blade is flickering and failing.
Patient Sequence No: 1, Text Type: D, B5


[13044658] (b)(4). The complaint case sample was rec'd w/some outward damage noticeable to the plastic carrying case. One of the locking latches was missing/broken off the case. Upon opening the kit it was also noted that one of the focs light pipes had been removed from a blade and was laying loose in the case. Also, during visual inspection the handle did not have a battery pack installed, and there were no loose battery packs in the carrying case. Since the laryngoscope handle did not arrive w/ a battery pack installed, a known good kit battery pack (new design) was installed into the handle. Multiple focs blades from the kit were attached to the handle and engaged. All of the blade lights were energized w/no flickering, no interruptions. The complaint cannot be confirmed. No corrective/preventative actions assigned. No further action required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1044475-2015-00141
MDR Report Key4658477
Report Source06,07
Date Received2015-03-27
Date of Report2015-03-11
Date of Event2015-03-11
Date Mfgr Received2015-03-11
Date Added to Maude2015-05-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMARGIE BURTON, RN
Manufacturer Street3015 CARRINGTON MILL BLVD
Manufacturer CityMORRISVILLE NC 27560
Manufacturer CountryUS
Manufacturer Postal27560
Manufacturer Phone9194334965
Manufacturer G1TELEFLEX MEDICAL
Manufacturer Street2917 WECK DR.
Manufacturer CityRESEARCH TRIANGLE PARK NC 27709
Manufacturer CountryUS
Manufacturer Postal Code27709
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRUSCH MRI COND FIBER OPTIC LARYN KIT
Generic NameLARYNGOSCOPE KIT
Product CodeOGH
Date Received2015-03-27
Returned To Mfg2015-03-19
Catalog Number005852300
Lot Number103301
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX MEDICAL
Manufacturer AddressRESEARCH TRIANGLE PARK NC US


Patients

Patient NumberTreatmentOutcomeDate
10 2015-03-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.