CARBON TANK TIMER CONTROL HEAD

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-04-02 for CARBON TANK TIMER CONTROL HEAD manufactured by Mar Cor.

Event Text Entries

[19356151] On (b)(6) 2014 we had a major problem with our water pre-treatment equipment made by mar cor purification causing significant delays in pt treatments for hemodialysis. The timer control heads had failed on the carbon tanks which are used to remove chlorine from water used in hemodialysis. Mar cor was called in to remediate the problem and pts had to be run outside the normal treatment are on portable reverse osmosis units until the proper repairs were made. Upon closer inspection, the plunger valve on the carbon tank timer controller had corroded and was stuck causing the carbon tank not to backwash at night and regenerate. Mar cor had made a change to the part which was made of brass originally and now was made of plastic to help prevent corrosion to this critical piece.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5041950
MDR Report Key4667524
Date Received2015-04-02
Date of Report2015-04-02
Date of Event2014-12-11
Date Added to Maude2015-04-08
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameCARBON TANK TIMER CONTROL HEAD
Generic NameWATER PURIFICATION SYSTEM
Product CodeFIP
Date Received2015-04-02
OperatorHEALTH PROFESSIONAL
Device Sequence No1
Device Event Key0
ManufacturerMAR COR
Manufacturer AddressPLYMOUTH MN 55447 US 55447


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-04-02

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.