GUTTACORE OBTURATOR (PINK) GCP04026

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2015-04-02 for GUTTACORE OBTURATOR (PINK) GCP04026 manufactured by Dentsply Tulsa Dental Specialties.

Event Text Entries

[16457625] In this event, it was reported that a dentist overfilled a patient's canal with gutta-core; no intervention was required. A review of the dentist's technique showed that he did not have a clear understanding of the obturation technique employed. Dentsply reviewed the protocols for usage with the dentist.
Patient Sequence No: 1, Text Type: D, B5


[16695992] While no serious injury resulted in this event, there have been previous reports received where overfilling of a root canal resulted in a serious injury. Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such. As such, this event is reportable per 21 cfr part 803. The device was not returned for evaluation and the lot number was not provided for retained-product testing and/or dhr review.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2320721-2015-00004
MDR Report Key4669529
Report Source05
Date Received2015-04-02
Date of Report2015-03-04
Date of Event2015-03-03
Date Mfgr Received2015-03-04
Date Added to Maude2015-05-14
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactHELEN LEWIS
Manufacturer Street221 W. PHILADELPHIA ST. STE. 60
Manufacturer CityYORK PA 17401
Manufacturer CountryUS
Manufacturer Postal17401
Manufacturer Phone7178457511
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGUTTACORE OBTURATOR (PINK)
Generic NameGUTTA-PERCHA
Product CodeEKM
Date Received2015-04-02
Catalog NumberGCP04026
Lot NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerDENTSPLY TULSA DENTAL SPECIALTIES
Manufacturer AddressJOHNSON CITY TN US


Patients

Patient NumberTreatmentOutcomeDate
10 2015-04-02

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.