ACCURUS HIGH BRIGHTNESS ILLUMINATOR SYSTEM 8065750290

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,01 report with the FDA on 2015-04-08 for ACCURUS HIGH BRIGHTNESS ILLUMINATOR SYSTEM 8065750290 manufactured by Alcon Research, Ltd. - Irvine Technology Center.

Event Text Entries

[5694210] A customer reported that the light source shut itself down during a vitrectomy procedure. The system was rebooted and the procedure was completed with the same system and accessories. There was no harm to the patient.
Patient Sequence No: 1, Text Type: D, B5


[13225506] Investigation including root cause analysis is in progress. A supplemental mdr will be filed as necessary in accordance with 21 cfr 803. 56 when additional reportable information becomes available. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[16696041] The system was examined and the reported event was replicated. The system was tested and found to meet product specifications. The light source was manufactured and released on january 5, 2007. Based on qa assessment, the product met specifications at the time of release. The system was found to meet specifications. Therefore, the root cause of the reported event cannot be determined conclusively. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2028159-2015-05566
MDR Report Key4669730
Report Source00,01
Date Received2015-04-08
Date of Report2015-06-26
Date of Event2015-03-10
Date Mfgr Received2015-06-02
Date Added to Maude2015-05-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. EDDIE DARTON, MD, JD
Manufacturer StreetMAIL STOP AB2-6 MAIL STOP R3-48
Manufacturer CityFORT WORTH TX 76134
Manufacturer CountryUS
Manufacturer Postal76134
Manufacturer Phone8175686660
Manufacturer G1ALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
Manufacturer Street15800 ALTON PARKWAY
Manufacturer CityIRVINE CA 92618
Manufacturer CountryUS
Manufacturer Postal Code92618
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameACCURUS HIGH BRIGHTNESS ILLUMINATOR SYSTEM
Generic NameENDOILLUMINATOR
Product CodeMPA
Date Received2015-04-08
Model NumberNA
Catalog Number8065750290
Lot NumberASKU
ID Number1.18 - 2.18
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerALCON RESEARCH, LTD. - IRVINE TECHNOLOGY CENTER
Manufacturer Address15800 ALTON PARKWAY IRVINE CA 92618 US 92618


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-04-08

© 2026 FDA.report