MAUDE MDR 467087

MDR report key
467087
Report number
MW1028739
Event key
0
Event type
3
Date of event
2003-04-22
Date received
2003-06-16
Adverse event
3
Product problem
3
Patients in event
0
Reporter occupation
0
Health professional
3
Initial report to FDA
0
Event location
3

Manufacturer Contact#

Report source
P
Manufacturer link flag
N

Devices#

Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1VAXCEL WITH PASV PICC 4FR SLVASCULAR ACCESS DEVICEBOSTON SCIENTIFIC CORPFKO***NN

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12003-06-1601. O

Event Narratives#

D

Patient 1

ADD'L INFO REC'D FROM MFR 10/10/03: ALL OF THE REPORTED EXPERIENCES DESCRIBE A BREAK IN THE EXTENSION TUBE (OR A HOLE) THAT DEVELOPED DURING USE. THERE WERE NO PT INJURIES. NONE OF THE INCIDENTS WERE REPORTED TO BSC BY THE CUSTOMER. MFR ENTERED THE INCIDENTS INTO ITS COMPLAINT SYSTEM AND INITIATED FOLLOW-UP TO OBTAIN A DETAILED DESCRIPTION. RPTR STATED THAT ALL OF THE SAMPLE PRODUCTS INVOLVED HAD BEEN DISCARDED. THEY HAD NO ADD'L INFO ABOUT THE INCIDENTS, EXCEPT THAT THE TUBING FRACTURED AT THE JUNCTION OF THE EXTENSION PART OF THE CATHETER AND THE HUB. NO SAMPLE WAS RETURNED FOR METHODOLOGIES TO BE APPLIED. NO SAMPLE WAS RETURNED FOR IDENTIFICATION/CONFIRMATION OF THE FAILURE MODE(S). NO CONCLUSIONS COULD BE MADE, SINCE NO SAMPLES WERE RETURNED. ALTHOUGH BOSTON SCIENTIFIC CORP DID NOT RECEIVE A PRODUCT SAMPLE FROM THE FACILITY FOR THESE SPECIFIC MEDWATCH REPORTS, IT HAS COMPLAINTS ON FILE FOR SIMILAR EXPERIENCES WITH THIS DEVICE. CO'S INVESTIGATIONS INTO THESE COMPLAINTS HAD ALREADY INDICATED THAT INAPPROPRIATE USE OF ALCOHOL OR ACETONE CAN RESULT IN THIS TYPE OF FAILURE. CO INSTITUTED INSERVICE TRAINING AMONG CUSTOMERS, INCLUDING THIS CUSTOMER TO REINFORCE THE PROPER HANDLING AND CARE PER THE DFU. THE DFU CONTAINS A PRECAUTION STATING "CARE SHOULD BE TAKEN WHEN USING ALCOHOL OR ACETONE. ALLOW FOR COMPLETE DRYING OF ALCOHOL OR ACETONE DURING ROUTINE CARE AND MAINTENANCE". NONE OF THE DEVICES INVOLVED IN THESE INCIDENTS WERE RETURNED TO BOSTON SCIENTIFIC. THE DEVICE WAS DESTROYED/DISCARDED BY THE CUSTOMER. ALTHOUGH THERE WAS NO CUSTOMER COMPLAINT SUBMITTED OR PRODUCT RETURNED FOR THESE MEDWATCH REPORTS, THIS CUSTOMER HAD RETURNED PRODUCT PREVIOUSLY AND THE INFO GAINED FROM EVALUATION OF THAT PRODUCT WAS USED IN THIS RESPONSE. DURING CO'S INVESTIGATIONS, IT CONFIRMED THE CATHETER EXTENSION TUBE'S SUSCEPTIBILITY TO KINKING AND FLEXING IN COMBINATION WITH ALCOHOL CONTACT. IN ADDITION, CO DISCOVERED THAT THE EXTENSION TUBE'S RESISTANCE TO KINKING AND ALCOHOL COULD BE MADE MORE ROBUST WITH A MINOR PROCESSING CHANGE. THE EXTENSION TUBES ARE THEN PROCESSED THROUGH THE ORIGINAL CATHETER ASSEMBLY PROCESS. THIS CHANGE WAS APPROVED AND IMPLEMENTED IN JUNE 2003.

D

Patient 1

VAXCEL WITH PASV PICC 4FR. SINGLE LUMEN BROKE AT THE JUNCTION OF THE HUB AND EXTERNAL PART OF CATHETER. DUE TO THIS BREAK ANOTHER VAXCEL (PICC LINE CHANGED OVER THE WIRE DONE).