OLYMPUS ENDOSONIC ULTRASONIC CLEANER ENDOSONIC CLEANER 7500617

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,06,07 report with the FDA on 2015-04-10 for OLYMPUS ENDOSONIC ULTRASONIC CLEANER ENDOSONIC CLEANER 7500617 manufactured by Keymed (medical & Industrial Equipment), Ltd..

Event Text Entries

[5710874] The olympus endosonic ultrasonic cleaner is a product which enables the cleaning of olympus endoscopic accessories and ancillary equipment by qualified technical or nursing staff in the cleaning room of a medical facility where flexible endoscopic accessories are reprocessed. The items to be cleaned are placed in a purpose designed basket and then immersed in the endosonic tank, which is filled to the appropriate level with ultrasonic cleaning solution. The fluid is agitated ultrasonically to remove debris and contamination from the items. Following cleaning, the items must be reprocessed in line with their instructions for use. Keymed has been made aware of an event where during the cleaning process, the facility staff smelled something burning from the inside of the endosonic cleaner. There is no report of injury to patient nor user and this report is submitted in an abundance of caution.
Patient Sequence No: 1, Text Type: D, B5


[13203118] An initial investigation has taken place at olympus (b)(4), which has confirmed that a transducer has become detached from the underside of the tank and that the adhesive which bonds the transducer to the tank has a burned appearance. Keymed are awaiting the return of the subject device. Once this happens a full investigation will take place.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9611174-2015-00003
MDR Report Key4693136
Report Source01,06,07
Date Received2015-04-10
Date of Report2015-03-16
Date of Event2015-02-26
Date Mfgr Received2015-03-16
Device Manufacturer Date2012-08-01
Date Added to Maude2015-05-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactALISON PRIOR, RA MANAGER
Manufacturer StreetJOURNEYMAN'S WAY, TEMPLE FARM INDUSTRIAL ESTATE
Manufacturer CitySOUTHEMD ON SEA, ESSEX SS2 5TF
Manufacturer CountryUK
Manufacturer PostalSS2 5TF
Manufacturer Phone1702616333
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS ENDOSONIC ULTRASONIC CLEANER
Generic NameENDOSONIC CLEANER
Product CodeFLG
Date Received2015-04-10
Model NumberENDOSONIC CLEANER
Catalog Number7500617
Lot Number21213284
OperatorOTHER
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerKEYMED (MEDICAL & INDUSTRIAL EQUIPMENT), LTD.
Manufacturer AddressSOUTHEND-ON-SEA UK


Patients

Patient NumberTreatmentOutcomeDate
10 2015-04-10