[19986883] The user facility reported their reliance 220 cart washer was emitting steam from around the doors during a cycle. The fire alarm was triggered due to the reported event, but no evacuations occurred. No report of injury or procedural delay or cancellation. The fire department was not dispatched to the user facility.
Patient Sequence No: 1, Text Type: D, B5


[20430657] A steris field service technician arrived onsite, inspected the unit, and identified the drying fan's belts required replacement. Due to the damaged belt on the drying fan, the fan was not able to clear the steam and condense it from within the cart washer effectively and thus caused the reported event. The technician replaced the drying fan belt. He tested the unit, ensured the steam cleared properly, and confirmed it to be operating according to specification. This unit is not under steris contract agreement for maintenance.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9680353-2015-00030
MDR Report Key4694944
Report Source06
Date Received2015-04-15
Date of Report2015-04-15
Date of Event2015-03-17
Date Mfgr Received2015-03-17
Date Added to Maude2015-05-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS KATHRYN CADORETTE
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927231
Manufacturer G1STERIS CANADA CORPORATION
Manufacturer Street490, ARMAND-PARIS
Manufacturer CityQUEBEC, QUEBEC GIC8A3
Manufacturer CountryCA
Manufacturer Postal CodeGIC 8A3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRELIANCE 220 CART WASHER
Generic NameWASHER
Product CodeFLH
Date Received2015-04-15
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS CANADA CORPORATION
Manufacturer Address490 ARMAND-PARIS QUEBEC, QUEBEC GIC8A3 CA GIC 8A3


Patients

Patient NumberTreatmentOutcomeDate
10 2015-04-15