MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07 report with the FDA on 2003-06-27 for ARTICULATING ARM FOR CEILING MOUNT SYSTEM 5787 manufactured by E-z-em, Inc..
[19155143]
Customer reported that the percupump 1a injector ceiling mount accessory broke causing the entire system to crash to the floor. It appeared that the ceiling mount had a stress fracture and finally broke near the first connecting joint from the mount base attached to the ceiling. The injector system was removed and replaced with a new empower ct demo unit. No one was injured as a result of the event.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2411512-2003-00008 |
MDR Report Key | 469577 |
Report Source | 05,06,07 |
Date Received | 2003-06-27 |
Date of Report | 2003-05-28 |
Date of Event | 2003-05-28 |
Date Facility Aware | 2003-05-28 |
Report Date | 2003-05-28 |
Date Reported to Mfgr | 2003-05-28 |
Date Added to Maude | 2003-07-07 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 0 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | GINA VOGELSBERG, MGR |
Manufacturer Street | 717 MAIN STREET |
Manufacturer City | WESTBURY NY 11590 |
Manufacturer Country | US |
Manufacturer Postal | 11590 |
Manufacturer Phone | 8005444624 |
Manufacturer G1 | E-Z-EM, INC. |
Manufacturer Street | 717 MAIN STREET |
Manufacturer City | WESTBURY NY 11590 |
Manufacturer Country | US |
Manufacturer Postal Code | 11590 |
Single Use | 3 |
Remedial Action | RL |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ARTICULATING ARM FOR CEILING MOUNT SYSTEM |
Generic Name | CT INJECTOR ACCESSORY |
Product Code | FIH |
Date Received | 2003-06-27 |
Model Number | NA |
Catalog Number | 5787 |
Lot Number | UNK |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | N |
Device Age | UNKNOWN |
Device Eval'ed by Mfgr | R |
Implant Flag | N |
Date Removed | A |
Device Sequence No | 1 |
Device Event Key | 458445 |
Manufacturer | E-Z-EM, INC. |
Manufacturer Address | 717 MAIN ST. WESTBURY NY 11590 US |
Baseline Brand Name | ARTICULATING ARM FOR CEILING MOUNT SYSTEM |
Baseline Generic Name | CT INJECTOR ACCESSORY |
Baseline Model No | NA |
Baseline Catalog No | 5787 |
Baseline ID | NA |
Baseline Device Family | ACCESSORY FOR PERCUPUMP INJECTOR SYSTEM |
Baseline Shelf Life [Months] | NA |
Baseline PMA Flag | N |
Baseline 510K PMN | Y |
Premarket Notification | K864227 |
Baseline Preamendment | N |
Baseline Transitional | N |
510k Exempt | N |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|