INGENUITY TF PET/MR 882380

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07,company representati report with the FDA on 2015-04-14 for INGENUITY TF PET/MR 882380 manufactured by Philips Medical Systems(cleveland), Inc..

Event Text Entries

[5638853] With the info provided, the customer reported that after performing a clinical pet/mr brain scan, the 256x256 (2mm) and the 576x576 (4mm) attenuation maps (mu-map) data was misaligned. The 4mm attenuation map was shifted to the left and up about 1 voxel (2 to 4 mm). This resulted in the reconstructions to have a left to right difference (profile ramp) in pet activity concentration and suv valve. The clinical product specialist stated that there has been no report of harm or image misinterpretation to a pt as a result of this issue.
Patient Sequence No: 1, Text Type: D, B5


[13134144] (b)(4). We will file a follow up mdr at the completion of the investigation. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[32488253] (b)(4). Initial mdr mailed on april 14, 2015. On (b)(6) 2015, the customer reported that after performing a clinical pet/mr brain scan, the 256x256 (2mm) and the 576x576 (4mm) attenuation maps (mu-map) data was misaligned. The 4mm attenuation map was shifted to the left and up about 1 voxel (2 to 4 mm). This resulted in the reconstructions to have a left to right difference (profile ramp) in pet activity concentration and suv value. The clinical product specialist stated that there has been no report of harm or image misinterpretation to a patient as a result of this issue. Further review of the information provided by the customer has determined that the variance is within the acceptable tolerance as defined in the following mitigation: gemsrs4158: after software calibration, the maximum error between the position of a point in the mr slice and the position of the corresponding point in the pet field of view shall be less than 4 mm. The observed variance was within the 4mm tolerance and therefore no correction was applicable for this issue. No further cause determination is applicable because further investigation of the information provided by the customer has determined that the variance is within the acceptable tolerance as defined in gemsrs4158: after software calibration, the maximum error between the position of a point in the mr slice and the position of the corresponding point in the pet field of view shall be less than 4 mm.
Patient Sequence No: 1, Text Type: N, H10


[32488255] With the information provided, the customer reported that after performing a clinical pet/mr brain scan, the 256x256 (2mm) and the 576x576 (4mm) attenuation maps (mu-map) data was misaligned. The 4mm attenuation map was shifted to the left and up about 1 voxel (2 to 4 mm). This resulted in the reconstructions to have a left to right difference (profile ramp) in pet activity concentration and suv value. The clinical product specialist stated that there has been no report of harm or image misinterpretation to a patient as a result of this issue.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1525965-2015-00101
MDR Report Key4698412
Report Source05,06,07,COMPANY REPRESENTATI
Date Received2015-04-14
Date of Report2015-03-16
Date Mfgr Received2015-03-16
Date Added to Maude2015-05-28
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMR. DAN BROWN
Manufacturer Street595 MINER RD
Manufacturer CityCLEVELAND OH 44143
Manufacturer CountryUS
Manufacturer Postal44143
Manufacturer Phone4404833000
Manufacturer G1PHILLIPS MEDICAL SYSTEMS
Manufacturer Street595 MINER RD MR TRAINING DEPARTMENT
Manufacturer CityCLEVELAND OH 44143
Manufacturer CountryUS
Manufacturer Postal Code44143
Single Use3
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report3

Device Details

Brand NameINGENUITY TF PET/MR
Generic NameTOMOGRAPHIC IMAGER COMBINING EMISSION CT WITH NUCLEAR MR
Product CodeOUO
Date Received2015-04-14
Model Number882380
Catalog Number882380
ID NumberN/A
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPHILIPS MEDICAL SYSTEMS(CLEVELAND), INC.
Manufacturer Address595 MINER RD. CLEVELAND OH 44143 US 44143


Patients

Patient NumberTreatmentOutcomeDate
10 2015-04-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.