OLYMPUS TROCARS WA58365S

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07 report with the FDA on 2015-04-17 for OLYMPUS TROCARS WA58365S manufactured by Olympus Winter & Ibe Gmbh.

Event Text Entries

[5737598] Olympus was informed that during a video assisted thoracoscopic surgery (vats) on the left upper lobectomy procedure, the tip of the trocar spike became fractured and broke off inside the patient's chest cavity. The device fragment was retrieved using a sponge stick. However, the user facility was uncertain if all of the device fragments were retrieved from the patient. No x-ray was performed. There was no patient injury reported. Additionally, the procedure was successfully completed using a different but similar device. The patient was reported to be in good condition.
Patient Sequence No: 1, Text Type: D, B5


[13235776] The device referenced in this report has not been returned to olympus for evaluation. The exact cause of the reported event could not be conclusively determined at this time. If additional information or if the device is received at a later time, this report will be supplemented.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2951238-2015-00173
MDR Report Key4712020
Report Source05,06,07
Date Received2015-04-17
Date of Report2015-04-17
Date of Event2015-03-31
Date Mfgr Received2015-04-01
Date Added to Maude2015-06-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactNOEMI SCHAMBACH
Manufacturer Street2400 RINGWOOD AVE.
Manufacturer CitySAN JOSE CA 95131
Manufacturer CountryUS
Manufacturer Postal95131
Manufacturer Phone4089355002
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameOLYMPUS TROCARS
Generic NameTROCARS SPIKE
Product CodeEWY
Date Received2015-04-17
Model NumberWA58365S
Catalog NumberWA58365S
Lot Number087W-0014
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerOLYMPUS WINTER & IBE GMBH
Manufacturer AddressKUEHNSTRASSE 61 HAMBURG 22045 GM 22045


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-04-17