3613 LITE GLOVE 31140216

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07 report with the FDA on 2015-04-21 for 3613 LITE GLOVE 31140216 manufactured by Covidien.

Event Text Entries

[5736107] Customer states they opened the package and the light glove was sliced.
Patient Sequence No: 1, Text Type: D, B5


[13272911] An investigation is currently underway. Upon completion, the results will be forwarded.
Patient Sequence No: 1, Text Type: N, H10


[25372836] The device history record was reviewed and indicated that the product was released accomplishing all quality standards. A sample was received for evaluation. A visual inspection was performed and the reported issue was confirmed. The handle of the lite glove was ripped. The split in the lite glove handle is caused by a pleat generated during the thermoforming process due to geometry of the product/ mold design. This pleat creates weakness on the neck of the lite glove. A formal corrective and preventative action was opened for this reported issue. The manufacturing personnel were trained and made aware of this reported issue. If additional information is received warranting further analysis, the investigation will be resumed. This complaint will be used for tracking and trending purposes.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9612030-2015-00046
MDR Report Key4715036
Report Source05,06,07
Date Received2015-04-21
Date of Report2015-04-16
Date of Event2015-04-10
Date Mfgr Received2015-08-04
Date Added to Maude2015-05-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactTHOM MCNAMARA
Manufacturer Street15 HAMPSHIRE STREET
Manufacturer CityMANSFIELD MA 02048
Manufacturer CountryUS
Manufacturer Postal02048
Manufacturer Phone5084524811
Manufacturer G1COVIDIEN
Manufacturer StreetCALLE 9 SUR NO. 125 CUIDAD INDUSTRIAL
Manufacturer CityTIJUANA
Manufacturer CountryMX
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name3613 LITE GLOVE
Generic NameLITE GLOVE
Product CodeFQP
Date Received2015-04-21
Model Number31140216
Catalog Number31140216
Lot Number5002108064X
OperatorOTHER
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerCOVIDIEN
Manufacturer AddressCALLE 9 SUR NO. 125 CUIDAD INDUSTRIAL TIJUANA MX


Patients

Patient NumberTreatmentOutcomeDate
10 2015-04-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.