TEST THROMBIN REAGENT OWHM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2015-04-28 for TEST THROMBIN REAGENT OWHM manufactured by Siemens Healthcare Diagnostics Products Gmbh.

Event Text Entries

[5705786] A customer running a lab for remote samples was contacted by one of his clients that a surgery had been postponed based on a potentially falsely prolonged thrombin time (tt) result. The patient result was reported to the physician. It is unknown if patient treatment was otherwise altered or prescribed on the basis of the potentially falsely prolonged thrombin time (tt) result. There is no report of adverse outcome to the patients as a result of the potentially falsely prolonged thrombin time (tt) result.
Patient Sequence No: 1, Text Type: D, B5


[13228737] The cause of the discrepant potentially falsely prolonged thrombin time (tt) result is unknown. The affected sample material was not available for internal investigation. However, investigation with other samples is ongoing to assess a potential for prolonged tt results in sec. Based on qc and the device function with other assays and parameters, there is no indication for a potential performance issue with the sysmex ca-7000 instrument. The instrument is performing within specifications.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9610806-2015-00009
MDR Report Key4732547
Report Source01,05,06
Date Received2015-04-28
Date of Report2015-03-30
Date of Event2015-03-27
Date Mfgr Received2015-03-30
Device Manufacturer Date2014-07-18
Date Added to Maude2015-05-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationMEDICAL TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJAMES MORGERA
Manufacturer StreetGLASGOW BUSINESS COMMUNITY PO BOX 6101
Manufacturer CityNEWARK DE 197146101
Manufacturer CountryUS
Manufacturer Postal197146101
Manufacturer Phone3026318356
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer Street76 EMIL VON BEHRING STRASSE
Manufacturer CityMARBURG, 35041
Manufacturer CountryGM
Manufacturer Postal Code35041
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTEST THROMBIN REAGENT
Generic NameTEST THROMBIN REAGENT
Product CodeGJA
Date Received2015-04-28
Catalog NumberOWHM
Lot Number44083
Device Expiration Date2016-06-12
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS PRODUCTS GMBH
Manufacturer Address76 EMIL VON BEHRING STRASSE MARBURG, 35041 GM 35041


Patients

Patient NumberTreatmentOutcomeDate
10 2015-04-28

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.