Gigly saw broke while cutting through mandible. All pieces retreived. No harm to patient. Physician reported this event when using this instrument. According to clinical engineering report, after repeated uses, the spring steel had become brittle and broken when it was flexedinvalid data - regarding single use labeling of device. Patient medical status prior to event: unknown. Invalid data - regarding multiple patient involvement. Invalid data - on device service/maintenance. No data - regarding date last serviced. Service provided by: invalid data. Invalid data - service records availability. Invalid data - regarding whether event presents imminent hazard. Invalid data - whether device used as labeled/intended. Device was evaluated after the event. Method of evaluation: actual device involved in incident was evaluated, visual examination. Results of evaluation: end of life - expected. Conclusion: device failure directly caused event. Certainty of device as cause of or contributor to event: yes. Corrective actions: device permanently removed from service. Invalid data - on device destroyed/disposed of status.
Patient Sequence No: 1, Text Type: D, B5