GIGLY SAW

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1992-07-31 for GIGLY SAW manufactured by Unknown.

Event Text Entries

[2705] Gigly saw broke while cutting through mandible. All pieces retreived. No harm to patient. Physician reported this event when using this instrument. According to clinical engineering report, after repeated uses, the spring steel had become brittle and broken when it was flexedinvalid data - regarding single use labeling of device. Patient medical status prior to event: unknown. Invalid data - regarding multiple patient involvement. Invalid data - on device service/maintenance. No data - regarding date last serviced. Service provided by: invalid data. Invalid data - service records availability. Invalid data - regarding whether event presents imminent hazard. Invalid data - whether device used as labeled/intended. Device was evaluated after the event. Method of evaluation: actual device involved in incident was evaluated, visual examination. Results of evaluation: end of life - expected. Conclusion: device failure directly caused event. Certainty of device as cause of or contributor to event: yes. Corrective actions: device permanently removed from service. Invalid data - on device destroyed/disposed of status.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number4734
MDR Report Key4734
Date Received1992-07-31
Date of Report1992-04-13
Date of Event1992-04-02
Date Facility Aware1992-04-02
Report Date1992-04-13
Date Reported to Mfgr1992-04-13
Date Added to Maude1993-06-07
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameGIGLY SAW
Product CodeDZH
Date Received1992-07-31
ID NumberNONE
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AvailabilityY
Implant Flag*
Device Sequence No1
Device Event Key4450
ManufacturerUNKNOWN


Patients

Patient NumberTreatmentOutcomeDate
101. Other 1992-07-31

© 2026 FDA.report