IVORY CLAMP IVORY CLAMP SS 12A 50057348

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05 report with the FDA on 2015-04-29 for IVORY CLAMP IVORY CLAMP SS 12A 50057348 manufactured by Heraeus Kulzer, Llc.

Event Text Entries

[5826754] Dental assistant reports an ivory clamp ss 12a reg molar broke in a patient's mouth today. The clamp broke into two pieces and both pieces were caught by the rubber dam material. When asked, the da said they were using hu-friedy rubber dam forceps. I recommended in the future that they use only genuine ivory rubber dam forceps which are stamped "made in usa", as these forceps are the ones manufactured to specifications of the clamp extension. Other forceps may allow the prongs to open too far and hyper extend the wings of the clamp, potentially leading to fracture. Agreed to replace ivory clamp ss 12a reg molar.
Patient Sequence No: 1, Text Type: D, B5


[13323575] Although we have not established that the device caused or contributed to the event, we're reporting it to be compliant with 21 cfr part 803 and out of an abundance of caution. The rubber dam clamp has been thrown away by the dental staff.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1821514-2015-00018
MDR Report Key4736161
Report Source01,05
Date Received2015-04-29
Date of Report2015-04-14
Date of Event2015-04-14
Date Facility Aware2015-04-14
Date Mfgr Received2015-04-14
Device Manufacturer Date2013-02-28
Date Added to Maude2015-06-01
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationDENTAL ASSISTANT
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. RITA ROGERS
Manufacturer Street300 HERAEUS WAY
Manufacturer CitySOUTH BEND IN 46614
Manufacturer CountryUS
Manufacturer Postal46614
Manufacturer Phone5742995409
Manufacturer G1HERAEUS KULZER, LLC
Manufacturer Street300 HERAEUS WAY
Manufacturer CitySOUTH BEND IN 46614251
Manufacturer CountryUS
Manufacturer Postal Code46614 2517
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIVORY CLAMP
Generic NameCLAMP, RUBBER DAM
Product CodeCEK
Date Received2015-04-29
Model NumberIVORY CLAMP SS 12A
Catalog Number50057348
Lot NumberI3
Device Expiration Date2015-02-28
OperatorDENTAL ASSISTANT
Device AvailabilityN
Device Age26 MO
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerHERAEUS KULZER, LLC
Manufacturer Address300 HERAEUS WAY SOUTH BEND IN 46614251 US 46614 2517


Patients

Patient NumberTreatmentOutcomeDate
10 2015-04-29