GEBAUER'S ETHYL CHLORIDE MIST SPRAY 0386-0001-02

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2015-04-28 for GEBAUER'S ETHYL CHLORIDE MIST SPRAY 0386-0001-02 manufactured by Gebauer Co..

Event Text Entries

[18544010] The pt was receiving dialysis and gebauer's ethyl chloride mist spray (lot 3068) was applied for pre-injection anesthesia. The pt reported burning, blistering, scabbing and discoloration after application of the product.
Patient Sequence No: 1, Text Type: D, B5


[18953364] The complaint as reported identified that the pt felt a burning sensation immediately after the product was applied that resulted in blistering, scabbing and discoloration. These adverse events are typically associated with frostbite that could result from over-application of the device. The complainant, who did not speak to the pt directly was told the product was applied by a nurse for approx 5 seconds. A 5 second spray is within the ifu but if the skin had turned white prior to the 5 seconds, the spray should have been stopped per the directions for use. Also, although rare, the adverse event could be the result of an allergic reaction to ethyl chloride or urticaria. Gebauer was told the pt required medical attention but no info was provided as to wat interventions were taken. Multiple attempts were made to obtain the unit back as well as seek further info on the injury and the medical attention required but we were informed the product would not be returned and no further info was available.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1519179-2015-00001
MDR Report Key4736168
Report Source05
Date Received2015-04-28
Date of Report2015-04-27
Date of Event2015-03-27
Date Mfgr Received2015-03-27
Device Manufacturer Date2014-04-29
Date Added to Maude2015-04-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactAMY PAUKPVITS
Manufacturer Street4444 E. 153RD ST.
Manufacturer CityCLEVELAND OH 44128
Manufacturer CountryUS
Manufacturer Postal44128
Manufacturer Phone2165813030
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGEBAUER'S ETHYL CHLORIDE MIST SPRAY
Generic NameCAPOCOOLANT
Product CodeMLY
Date Received2015-04-28
Model Number0386-0001-02
Lot Number3068
Device Expiration Date2016-04-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerGEBAUER CO.
Manufacturer AddressCLEVELAND OH US


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2015-04-28

© 2024 FDA.report