MARUS DENTAL CHAIR DC1320

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2015-05-01 for MARUS DENTAL CHAIR DC1320 manufactured by Marus.

Event Text Entries

[5827469] A dental professional was performing routine medical treatment to a patient when the swivel weld assembly broke on a marus dental chair causing the upper structure of the chair to fall off the lower section of the chair. There were no injuries reported.
Patient Sequence No: 1, Text Type: D, B5


[13322522] Upon evaluation by the local marus distributor technician, it was identified that the swivel weld assembly broke apart at the weld. The part was not given to the distributor by the doctor therefore ther part was not returned to marus for evaluation. The dental chair is over 13 years old and is past the life of the device.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1017522-2015-00006
MDR Report Key4745076
Report Source08
Date Received2015-05-01
Date of Report2015-04-03
Date of Event2015-04-02
Date Mfgr Received2015-04-03
Device Manufacturer Date2001-09-01
Date Added to Maude2015-06-11
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactFRANK RAY
Manufacturer Street11727 FRUEHAUF DR.
Manufacturer CityCHARLOTTE NC 28273
Manufacturer CountryUS
Manufacturer Postal28273
Manufacturer Phone7045877227
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMARUS DENTAL CHAIR
Generic NameDENTAL CHAIR
Product CodeEAZ
Date Received2015-05-01
Model NumberDC1320
Catalog NumberDC1320
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerMARUS
Manufacturer Address11727 FRUEHAUF DR. CHARLOTTE NC 28273 US 28273


Patients

Patient NumberTreatmentOutcomeDate
10 2015-05-01

© 2026 FDA.report