PRISM 6 CHANNEL ANALYZER 06A36-04

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05 report with the FDA on 2015-05-04 for PRISM 6 CHANNEL ANALYZER 06A36-04 manufactured by Abbott Manufacturing Inc.

Event Text Entries

[5887127] The customer stated that (b)(6) prism anti-hcv results were generated. An initial (b)(6) prism hcv result of (b)(6) was generated. The sample was repeated in duplicate on the same analyzer and (b)(6) results of (b)(6) and (b)(6) were generated. (b)(6) results were generated with biorad eia hcv, innolea hcv and viral load testing. There was no report of impact to patient management.
Patient Sequence No: 1, Text Type: D, B5


[13264699] (b)(4). A follow-up report will be submitted when the evaluation is complete.
Patient Sequence No: 1, Text Type: N, H10


[35078923] The customer observed (b)(6) prism hcv results with lot 44562li00 on the prism analyzer using reaction tray lot 43289m500. Testing was performed across three runs (two instruments) using one replicate of negative calibrator at the end of each run. All negative calibrator replicates were non-reactive, thus indicating the product is performing as expected. A malfunction was not identified as the issue occurred for a single patient, and the abbott prism hcv assay has a stated specificity claim of greater than or equal to 99. 73%. Customer complaint data was reviewed and no related adverse trends and no non-statistical trends were identified for the complaint issue. A review for non-conformances and deviations was performed and none were found. The prism hcv reagent package insert and the prism operations manual were reviewed and were found to adequately address the issue. The investigation did not identify a malfunction / deficiency.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1628664-2015-00145
MDR Report Key4746792
Report Source01,05
Date Received2015-05-04
Date of Report2015-04-10
Date of Event2015-04-10
Date Mfgr Received2015-05-28
Date Added to Maude2015-05-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactNOEMI ROMERO-KONDOS, RN BSN
Manufacturer Street100 ABBOTT PARK ROAD DEPT. 09B9, LCCP1-3
Manufacturer CityABBOTT PARK IL 600643537
Manufacturer CountryUS
Manufacturer Postal600643537
Manufacturer Phone224667-512
Manufacturer G1ABBOTT MANUFACTURING INC
Manufacturer Street1921 HURD DRIVE
Manufacturer CityIRVING TX 75038
Manufacturer CountryUS
Manufacturer Postal Code75038
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePRISM 6 CHANNEL ANALYZER
Generic NameAUTOMATED IMMUNOASSAY ANALYZER
Product CodeMZP
Date Received2015-05-04
Catalog Number06A36-04
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MANUFACTURING INC