TUBING, INSUFFLATION W/ .3 MIC FILT/ STRYKER ADA 28-0211

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2015-05-07 for TUBING, INSUFFLATION W/ .3 MIC FILT/ STRYKER ADA 28-0211 manufactured by S.l.r. Km7, Autopista Joaquin Balaguer.

Event Text Entries

[17589599] Customer reported an alarm that indicated there was a leak in the line. When the customer was contacted for additional information, it was reported that the "tube was obstructed at the filter".
Patient Sequence No: 1, Text Type: D, B5


[17735197] This report is being filed due to a retrospective review of complaints. This complaint has been re-opened for further investigation. A supplemental report will be filed if further investigation provides information which changes the content of this report. The original root cause was determined to be migration of the plasticizer into the filter housing. It was determined to change the filter housing. It was determined to change the filter material to a polycarbonate material which is stronger, higher impact, and more chemical resistant. It is also less affected by temperature.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3004605321-2015-00007
MDR Report Key4759440
Report Source06
Date Received2015-05-07
Date of Report2015-04-30
Date of Event2012-02-16
Date Facility Aware2012-02-16
Report Date2015-04-30
Date Mfgr Received2012-02-16
Device Manufacturer Date2011-11-07
Date Added to Maude2015-06-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street200 DEBUSK LANE
Manufacturer CityPOWELL TN 37849
Manufacturer CountryUS
Manufacturer Postal37849
Manufacturer Phone8653626157
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction NumberZ-2113-2013
Event Type3
Type of Report3

Device Details

Brand NameTUBING, INSUFFLATION W/ .3 MIC FILT/ STRYKER ADA
Generic NameINSUFFLATION TUBING, TBING W/ FILTER
Product CodeNKC
Date Received2015-05-07
Returned To Mfg2012-02-27
Catalog Number28-0211
Lot Number25962556
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerS.L.R. KM7, AUTOPISTA JOAQUIN BALAGUER
Manufacturer AddressPISANO FREE ZONE, BLDG. 49 SANTIAGO DR


Patients

Patient NumberTreatmentOutcomeDate
10 2015-05-07