FMS REF 4114

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2003-07-28 for FMS REF 4114 manufactured by Fms (future Medical Systems).

Event Text Entries

[17104909] Pt was undergoing a right arthroscopic debridement. Following the procedure, 5 to 6 small burns covering about a 4-inch by 4-inch area on the left inner medial thigh were noted. Burns are 2nd degree to full thickness that the plastic surgeon believes will heal with minimal scarring. The size and shape of the burns is consistent with the end of a metal rod that is used to immobilize the operative knee during the procedure. It was noted during the procedure that the pt "jumped" each time the shaver was activated. Initial investigation has revealed high micro leakage readings (. 987 ohms and 1181 ma) at future medical systems (fms)/stryker interface connector. (this interfaces the stryker 12k shaver with the fms irrigation/distention system). It was noted that the power cord was not fully seated, and when fully seated, the leakage and ground leakage dropped to. 235 ohms and 250 ma. Stryker has evaluated their equipment on site and found no problems.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number476197
MDR Report Key476197
Date Received2003-07-28
Date of Report2002-10-01
Date of Event2000-09-01
Date Added to Maude2003-08-08
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameFMS
Generic NameSHAVER INTERFACE
Product CodeHRX
Date Received2003-07-28
Model NumberREF 4114
Catalog NumberREF 4114
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age2 YR
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key465023
ManufacturerFMS (FUTURE MEDICAL SYSTEMS)
Manufacturer Address504 MCCORMICK DRIVE; SUITE T GLEN BERNIE MD 21061 US

Device Sequence Number: 2

Brand NameSTRYKER
Generic NameSHAVER HANDPIECE
Product CodeHRX
Date Received2003-07-28
Model Number275-701-500
Catalog Number12K SHAVER
Lot Number*
ID Number*
Device AvailabilityY
Device Age2 YR
Implant FlagN
Date Removed*
Device Sequence No2
Device Event Key486307
ManufacturerSTRYKER ENDOSCOPY
Manufacturer Address5900 OPTICAL CT SAN JOSE CA 95138 US

Device Sequence Number: 3

Brand NameFMS (FUTURE MEDICAL SYSTEMS)
Generic NameDUAL PUMP SYSTEM
Product CodeHRX
Date Received2003-07-28
Model NumberFMS DUO
Catalog NumberREF 5100-1
Lot Number*
ID Number35631380
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age2 YR
Implant FlagN
Date Removed*
Device Sequence No3
Device Event Key486308
ManufacturerFMS (FUTURE MEDICAL SYSTEMS)
Manufacturer Address504 MCCORMICK DRIVE SUITE T GLEN BERNIE MD 21061 US

Device Sequence Number: 4

Brand NameTRINITRON
Generic NameCOLOR VIDEO MONITOR
Product CodeFWB
Date Received2003-07-28
Model NumberPVM-20M2MDU
Catalog Number*
Lot Number*
ID Number1519159
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age2 YR
Implant FlagN
Date Removed*
Device Sequence No4
Device Event Key486310
ManufacturerSONT ELECTRONICS CORP
Manufacturer Address680 KINDERMACK RD ORAVELL NJ 07643 US

Device Sequence Number: 5

Brand NameSTRYKER
Generic NameLIGHT SOURCE
Product CodeFSY
Date Received2003-07-28
Model NumberX-6000
Catalog Number*
Lot Number*
ID Number35631211
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age2 YR
Implant FlagN
Date Removed*
Device Sequence No5
Device Event Key486311
ManufacturerSTRYKER ENDOSCOPY
Manufacturer Address5900 OPTICAL COURT SAN JOSE CA 95051 US

Device Sequence Number: 6

Brand NameTPS (TOTAL PERFORMANCE SYSTEM)
Generic NameSHAVER CONSOLE
Product CodeHRX
Date Received2003-07-28
Model NumberTPS CONSOLE
Catalog Number*
Lot Number*
ID Number35631206
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age2 YR
Implant FlagN
Date Removed*
Device Sequence No6
Device Event Key486312
ManufacturerSTRYKER INSTRUMENTS
Manufacturer Address5900 OPTICAL CT SAN JOSE CA 95138 US

Device Sequence Number: 7

Brand NameSTRYKER
Generic NameDIGITAL CAMERA UNIT (3 CHIP)
Product CodeFXM
Date Received2003-07-28
Model Number*
Catalog Number*
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age2 YR
Implant FlagN
Date Removed*
Device Sequence No7
Device Event Key486313
Manufacturer*
Manufacturer Address* * *


Patients

Patient NumberTreatmentOutcomeDate
101. Other; 2. Required No Informationntervention 2003-07-28

© 2026 FDA.report