FREE PSA 03289788190

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2015-05-13 for FREE PSA 03289788190 manufactured by Roche Diagnostics.

Event Text Entries

[5778201] The customer received questionable total psa total (free + complexed) psa - prostate-specific antigen (total psa) and free psa results for one patient sample from cobas e601 analyzer serial number (b)(4). The specific date of testing was not provided. The initial free-psa result was 1. 64 ng/ml and the initial total psa result was 0. 902 ng/ml. The customer recentrifuged the sample and repeated testing with the same results. (b)(4). Information concerning if any erroneous result was reported outside the laboratory was requested, but was not provided. No adverse event was reported.
Patient Sequence No: 1, Text Type: D, B5


[13446374] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[13488582] A specific root cause could not be identified as no patient sample could be provided for further investigation.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2015-03455
MDR Report Key4768797
Report Source01,05,06
Date Received2015-05-13
Date of Report2015-06-12
Date of Event2015-04-30
Date Mfgr Received2015-04-30
Date Added to Maude2015-05-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameFREE PSA
Generic NameTEST,PROSTATE SPECIFIC ANTIGEN,FREE,(NONCOMPLEXED) TO DISTINGUISH PROSTATE CA
Product CodeMTG
Date Received2015-05-13
Model NumberNA
Catalog Number03289788190
Lot Number178872
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2015-05-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.