TRIGLYCERIDES_2 REAGENT, CONCENTRATED 10697575

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06,foreign,health profe report with the FDA on 2015-05-14 for TRIGLYCERIDES_2 REAGENT, CONCENTRATED 10697575 manufactured by Siemens Healthcare Diagnostics Inc..

Event Text Entries

[5781782] The customer carried out a correlation study using patient samples, between the triglycerides_2 concentrated (trig_c) assay and the triglycerides assay on an advia chemistry 2400 instrument. The customer observed that there was a loss of linearity at around 2. 5-3. 0 mmol/l when using trig_c reagent lot 300275. It is unknown if any of the results obtained on the advia chemistry 2400 were reported to physician(s) there were no reports of adverse health consequences due to the results obtained for trig_c on the patient samples.
Patient Sequence No: 1, Text Type: D, B5


[13426766] Siemens has confirmed a low bias at the higher end of the triglyceride assay range on patient, quality control (qc), and linearity samples of advia chemistry trig_c reagent lot 300275 compared to alternate in date lots of trig_c, trig and trig_2 reagents. During siemens investigation, the trig_c reagent lot 300275 displayed a low bias that exceeded 25% at levels greater than approximately 425mg/dl for patients and quality controls. A positive (high) bias of approximately +10% was observed in qc samples at values less than 100mg/dl. The low bias occurred just prior to expiration of the affected kit lot 300275. Siemens diagnostics is investigating this issue.
Patient Sequence No: 1, Text Type: N, H10


[32158644] The initial mdr 2432235-2015-00232 was filed on may 14, 2015. Additional information (05/06/15): an actual root cause has not been established as the affected lot has expired. The current lots are being monitored for linearity on a monthly basis until the end of shelf life.
Patient Sequence No: 1, Text Type: N, H10


[32604090] The initial mdr 2432235-2015-00232 was filed on may 14, 2015. Supplemental mdr 2432235-2015-00232_s1 was filed on june 1, 2015. Additional information (07/13/15): the root cause of the low bias at the higher end of the triglyceride assay range on patient, quality control (qc), and linearity samples of advia chemistry trig_c reagent lot 300275 was determined to be a lower concentration of glycerokinase (gk) enzyme. The affected trig_c reagent lot 300275 expired on 12/31/2014. The customer has moved on to another lot that is unaffected.
Patient Sequence No: 1, Text Type: N, H10


[38903636] The initial mdr 2432235-2015-00232 was filed on may 14, 2015. Supplemental mdr 2432235-2015-00232_s1 was filed on june 1, 2015. Supplemental mdr 2432235-2015-00232_s2 was filed on august 12, 2015. Additional information (1/20/2016): during the process of monitoring all current lots siemens technical operations has confirmed that the trig_c assay reagent lot 338040 which was being monitored through end of shelf life is not linear to the instructions for use (ifu) claim of 550 mg/dl (62. 15 mmol/l). Siemens diagnostics is investigating this issue.
Patient Sequence No: 1, Text Type: N, H10


[41753866] Correction: supplemental mdr #3 states the date of awareness as 01/20/2016. The correct date is 01/13/2016. Additional information (2/29/2016): siemens could not conduct an investigation on trig_c reagent lot 300275 as it had expired. In date reagent lots 348297 and 359932 were monitored and based on internal studies siemens has confirmed that the two lots do not meet linearity claims as indicated in the instructions for use (ifu) of the assay. An urgent field safety notice (ufsn) chc16-03. A. Ous was sent to ous customers and a urgent medical device recall (umdr) chc16-03. A. Us was sent to us customers in march of 2016 to address reagent kit lots 348297 and 359932 only. Lot 300275 is expired and not part of this recall.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2432235-2015-00232
MDR Report Key4770751
Report Source01,05,06,FOREIGN,HEALTH PROFE
Date Received2015-05-14
Date of Report2016-01-20
Date of Event2014-12-19
Date Mfgr Received2016-02-29
Date Added to Maude2015-06-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactAARTI AZIZ
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242683
Manufacturer G1RANDOX LABORATORIES LTD. (REG# 8020890)
Manufacturer Street55 DIAMON ROAD CRUMLIN
Manufacturer CityCO. ANTRIM, BT294QY
Manufacturer CountryUK
Manufacturer Postal CodeBT29 4QY
Single Use3
Previous Use Code3
Removal Correction Number2432235-03/09/2016-001-R
Event Type3
Type of Report3

Device Details

Brand NameTRIGLYCERIDES_2 REAGENT, CONCENTRATED
Generic NameTRIGLYCERIDE_2 REAGENTS, CONCENTRATED
Product CodeCDT
Date Received2015-05-14
Model NumberTRIGLYCERIDES_2 REAGENT, CONCENTRATED
Catalog Number10697575
Lot Number300275
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address511 BENEDICT AVENUE TARRYTOWN NY 10591 US 10591


Patients

Patient NumberTreatmentOutcomeDate
10 2015-05-14

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.