IDRT, UNKNOWN XXX-IDRT

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,07,08,company represent report with the FDA on 2015-05-15 for IDRT, UNKNOWN XXX-IDRT manufactured by Integra Lifesciences Corp..

Event Text Entries

[17583252] It was reported that one week after implantation of the product, the wound became infected with pseudomonas aeruginosa. "the burns and plastics (units) are using the products not only for full thickness but for other wounds such as diabetic ulcers. " comments from the consultant plastic surgeon: "i don't know how things have got to this stage! As i explained over the phone, we have had a few integra pts become infected recently. The infections always developed after the initial post-op dressing change. All the pts were fine at first post-op dressing, indicating that there was no problem with the product or the operative technique used in theatre. The problem appears to be due to poor technique by certain nurses on the ward when carrying out subsequent dressing changes. I cannot give you the details requested as several pts are involved with different lots of integra used at different anatomical sites. In all cases to date the infections were treated successfully and the outcome was satisfactory without the need to remove the integra or to apply any further integra or other product. Apparently our so called link nurse reported the infections to a nurse in the infection control dept. From what i can gather, the latter contacted the mfr without first verifying what was going on with the med staff. None of the med staff, including both consultants, were ever contacted by the infection control dept. We were well aware of the problem and had already taken steps to remedy the situation by insisting with our nurses that strict sterile techniques were to be used when changing dressings on integra pts. I can assure you that to my knowledge there is no problem with the product and whether we like it or not we have to admit that there were serious shortcomings on the part of our ward nursing staff. "
Patient Sequence No: 1, Text Type: D, B5


[17740155] The device will not be returned since it remains implanted. Based on reported info, integra has initiated an investigation.
Patient Sequence No: 1, Text Type: N, H10


[50668260] Integra has completed their internal investigation on 06/04/2015. Results, dhr review: the reported product lot associated with this complaint for idrt - ts 8x10 is lot # 105nd0309885. This finished goods product lot stems from parent dispersion lot # 105n00306921. The manufacture date for this lot is 06/2014 and date of expiration is 06/2016. Overall, there were no issues identified during the batch record review that could be attributed to the failure mode of "infection" reported with the current complaint. This complaint is associated with an dirt-ts lot which is similar to idrt-sl and idrt-meshed products. There were approx (b)(4) of these products sold in the past year (april 2014 - april 2015). As per the trending query, there were (b)(4) similar complaints identified for infection. Therefore, the calculated complaint rate is (b)(4). Conclusion: the root cause is undetermined. As the product was not returned, the failure was unconfirmed. Additionally, there were no issues identified from the dhr review. Trending and risk analysis that would have attributed to this complaint. Per the complaint background and according to the surgeon, neither the product nor the operative technique caused the infections with the integra pts, as all pts were fine after the initial post-op dressing. The surgeon attributes the failure to the lack of sterile techniques used by the nursing ward at the hospital during subsequent post-op dressing changes for the involved pts. In all cases, the infections were treated successfully and the outcome was satisfactory without the need to remove the integra implant or to apply any further integra or other product.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1121308-2015-00009
MDR Report Key4780327
Report Source01,05,07,08,COMPANY REPRESENT
Date Received2015-05-15
Date of Report2015-04-20
Date Mfgr Received2015-04-20
Device Manufacturer Date2014-06-01
Date Added to Maude2015-05-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMARIA LEONARD
Manufacturer Street311 ENTERPRISE DR
Manufacturer CityPLAINSBORO NJ 08536
Manufacturer CountryUS
Manufacturer Postal08536
Manufacturer Phone6099362341
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIDRT, UNKNOWN
Generic NameINTEGRA
Product CodeMGR
Date Received2015-05-15
Catalog NumberXXX-IDRT
Lot Number105ND0309885
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerINTEGRA LIFESCIENCES CORP.
Manufacturer AddressPLAINSBORO NJ 08536 US 08536


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2015-05-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.