VITEK 2 AST-YS05 TEST KIT 411945

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2015-05-19 for VITEK 2 AST-YS05 TEST KIT 411945 manufactured by Biom?rieux, Inc..

Event Text Entries

[20755323] A customer in the united states contacted biom? Rieux to report a false susceptible fluconazole result (mic = 4) for a candida glabrata organism on the vitek? 2 ast-ys05 test kit (ref 411945, lot 285346710, expiry 11jun2016). The result was reported to the treating physician. The patient isolate was sent to a reference laboratory (arup) for testing. The reference laboratory obtained a resistant result (mic = 16) for fluconazole via broth micro-dilution (bmd). The physician claims treatment failure and delay to receive appropriate treatment >24 hours. Patient therapy was modified once the result from the reference laboratory was received. Biom? Rieux has requested culture submittal for internal testing.
Patient Sequence No: 1, Text Type: D, B5


[20882941] Device not received by manufacturer.
Patient Sequence No: 1, Text Type: N, H10


[34457464] Two candida glabrata isolates were submitted by the customer to biom? Rieux for evaluation. The organism identification of the isolates was confirmed (candida glabrata) and fluconazole testing included two test kits from the same lot of ast-ys05 as that tested by the customer and two test kits from a random lot. Also tested was broth microdilution (bmd), the reference method for fluconazole. Category calls for all testing was determined using the vitek? 2 breakpoints: <=8 s, 16-32 i, >=64 r. Isolate #1: susceptible mics = 2 were obtained from the four ast- ys05 test kits; bmd gave a susceptible mic = 1. Isolate #2: susceptible mics = 2 were obtained from the customer's lot of ast-ys05 cards and from the random lot, susceptible mics = 2 and 4 were obtained. Bmd gave a susceptible mic = 2. Based on the results of the investigation testing, the vitekc 2 ast-ys05 test kits are performing as expected and within performance claims. The customer's lack of correlation between vitekv 2 and bmd methods could not be duplicated. The 2008 clsi m27-a3 and the 2012 m27-s4, reference method for broth dilution antifungal susceptibility testing of yeast, contains a note which states "when an isolate is identified as candida glabrata and the mic is <=32, it should be considered if fluconazole is appropriate in the specific clinical context. "
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1950204-2015-00005
MDR Report Key4780357
Report Source05,06
Date Received2015-05-19
Date of Report2015-05-12
Date of Event2015-05-12
Date Mfgr Received2015-05-12
Device Manufacturer Date2014-12-11
Date Added to Maude2015-05-19
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. RYAN LEMELLE
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3147318582
Manufacturer G1BIOM
Manufacturer Street595 ANGLUM ROAD
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal Code63042
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameVITEK 2 AST-YS05 TEST KIT
Generic NameVITEK 2 AST-YS05 TEST KIT
Product CodeNGZ
Date Received2015-05-19
Catalog Number411945
Lot Number285346710
Device Expiration Date2016-06-11
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerBIOM?RIEUX, INC.
Manufacturer Address595 ANGLUM ROAD HAZELWOOD MO 63042 US 63042