MAUDE MDR 4787061

MDR report key
4787061
Report number
0001811755-2015-01829
Event key
0
Event type
3
Date of event
2015-04-23
Date received
2015-05-21
Adverse event
0
Product problem
1
Patients in event
0
Reporter occupation
0
Health professional
0
Initial report to FDA
0
Event location
0

Manufacturer Contact#

Report source
M
Manufacturer link flag
Y

Devices#

Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1LOANER TPS MICRODRIVERDRIVER, WIRE, AND BONE DRILL, MANUALSTRYKER INSTRUMENTS-KALAMAZOODZJ5100088000LY R
1368277LOANER TPS MICRODRIVERDRIVER, WIRE, AND BONE DRILL, MANUALSTRYKER INSTRUMENTS-KALAMAZOODZJ5100088000LR

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12015-05-210

Event Narratives#

D

Patient 1

THE LOANER TPS MICRODRIVER WAS RETURNED TO STRYKER INSTRUMENTS FOR SERVICE. DURING FUNCTIONAL TESTING BY A SERVICE TECHNICIAN, IT WAS FOUND THAT THE DEVICE WAS ABLE TO RUN IN SAFE MODE. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.

N

Patient 1

THE FAILURE FOR RUNNING IN SAFE MODE WAS CONFIRMED THROUGH FUNCTIONAL TESTING, DISASSEMBLY, AND VISUAL INSPECTION. THE TRIGGER ASSEMBLY WAS WORN. THE DEVICE WAS REPAIRED AND PLACED BACK INTO STRYKER STOCK.

D

Patient 1

THE LOANER TPS MICRODRIVER WAS RETURNED TO STRYKER INSTRUMENTS FOR SERVICE. DURING FUNCTIONAL TESTING BY A SERVICE TECHNICIAN, IT WAS FOUND THAT THE DEVICE WAS ABLE TO RUN IN SAFE MODE. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.