LABPRO COMPUTER N/A B1018-510

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2015-05-21 for LABPRO COMPUTER N/A B1018-510 manufactured by Beckman Coulter.

Event Text Entries

[18892930] It was reported that the labpro aegis computer used with walkaway instruments started smoking and produced a small amount of flames coming from the vent of the computer. A laboratory technician extinguished the flame and smothered the flame. No fire extinguisher was needed. Building facilities removed the computer from the lab. No power surges were observed prior to the event. It was reported that the panels already loaded onto the two walkaway instruments were read manually. There were no reports of injuries or adverse events as a result of this incident.
Patient Sequence No: 1, Text Type: D, B5


[19065670] Labpro computer is not a medical device and is used in conjunction with the walkaway instrument. The computer was returned for evaluation and it was determined, that dust accumulated due to the computer's electronic failure, likely started the smoke and resulted into the flame. The mother board failed and no additional testing could be performed. The manufacturer's field service engineer replaced the damaged computer and installed the required labpro software. The replacement computer was tested with the walkaway instruments and results met the performance specifications. Beckman coulter internal identified for this report is (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2919016-2015-00060
MDR Report Key4789565
Report Source05,06
Date Received2015-05-21
Date of Report2015-04-23
Date of Event2015-04-23
Date Mfgr Received2015-04-23
Date Added to Maude2015-06-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactNORMA RAHILL
Manufacturer Street1584 ENTERPRISE BLVD
Manufacturer CityWEST SACRAMENTO CA 95691
Manufacturer CountryUS
Manufacturer Postal95691
Manufacturer Phone9163742139
Manufacturer G1BECKMAN COULTER
Manufacturer Street2040 ENTERPRISE BLVD
Manufacturer CityWEST SACRAMENTO CA 95691
Manufacturer CountryUS
Manufacturer Postal Code95691
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLABPRO COMPUTER
Generic NameINSTRUMENT FOR AUTO READER & INTERPRETATION OF OVERNIGHT SUSCEPT. SYSTEMS
Product CodeLRG
Date Received2015-05-21
Returned To Mfg2015-04-29
Model NumberN/A
Catalog NumberB1018-510
Lot NumberN/A
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER
Manufacturer Address2040 ENTERPRISE BLVD WEST SACRAMENTO CA 95691 US 95691


Patients

Patient NumberTreatmentOutcomeDate
10 2015-05-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.