MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00 report with the FDA on 2015-05-22 for POWER SUPPLY 21 VISION ELECT FLAT MONITOR 0240030931 manufactured by Stryker Endoscopy-san Jose.
[5814943]
It was reported that one of the boom mounted monitors start smoking. The biomedical engineering team pulled the power brick out because it was burning & smoking.
Patient Sequence No: 1, Text Type: D, B5
[13404963]
Additional information will be provided once the investigation has been completed.
Patient Sequence No: 1, Text Type: N, H10
[29953525]
The reported device was not received for investigation; therefore the reported failure cannot be confirmed. The complaint will be closed without a detailed investigation report and based on probable root cause. In the event that the device is received, the complaint will be reopened, a full evaluation will be conducted, and the investigation will be updated with the new results. Probable root cause for the reported failure involving this device could have been caused by: display design, power supply material, isolation power failure. Fuses fail to break circuit due to excessive mains current leading to fire. The product was not returned for investigation therefore the reported failure mode was not confirmed. The failure mode will be monitored for future reoccurrence.
Patient Sequence No: 1, Text Type: N, H10
[29953526]
It was reported that one of the boom mounted monitors start smoking. The biomedical engineering team pulled the power brick out because it was burning & smoking.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 0002936485-2015-00411 |
MDR Report Key | 4792624 |
Report Source | 00 |
Date Received | 2015-05-22 |
Date of Report | 2015-04-28 |
Date of Event | 2015-04-28 |
Date Mfgr Received | 2015-04-28 |
Date Added to Maude | 2015-06-19 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | MR. THOMAS SHAFER |
Manufacturer Street | 5900 OPTICAL COURT |
Manufacturer City | SAN JOSE CA 95138 |
Manufacturer Country | US |
Manufacturer Postal | 95138 |
Manufacturer Phone | 4087542000 |
Manufacturer G1 | STRYKER ENDOSCOPY-SAN JOSE |
Manufacturer Street | 5900 OPTICAL COURT |
Manufacturer City | SAN JOSE CA 95138 |
Manufacturer Country | US |
Manufacturer Postal Code | 95138 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | POWER SUPPLY 21 VISION ELECT FLAT MONITOR |
Generic Name | CAMERA, TELEVISION, SURGICAL, WITHOUT AUDIO |
Product Code | FWB |
Date Received | 2015-05-22 |
Catalog Number | 0240030931 |
Operator | OTHER |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | R |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | STRYKER ENDOSCOPY-SAN JOSE |
Manufacturer Address | 5900 OPTICAL COURT SAN JOSE CA 95138 US 95138 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2015-05-22 |