VYGON VALUE LIFE PREP KIT AMS-8023CP-1

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-05-21 for VYGON VALUE LIFE PREP KIT AMS-8023CP-1 manufactured by Vygon Usa.

Event Text Entries

[5925163] Carefusion maxplus clear needless connector, ref #(b)(4), lot #15028062, exp 02/2020, unscrewed itself from the 10ml prefilled saline syringe that is contained in the vygon value life prep kit (kit for central line access) product code ams-8023cp-1 lot #14l23d, expiration 6/2016. This has happened several times with various patients with the combination of these two products. The maxplus clear needleless connector has not done this with refilled saline syringes that are not contained in this kit that are used in the facility. Diagnosis or reason for use: needless connector for vascular access.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW5042849
MDR Report Key4802298
Date Received2015-05-21
Date of Report2015-05-21
Date of Event2015-05-19
Date Added to Maude2015-05-29
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Sequence Number: 1

Brand NameVYGON VALUE LIFE PREP KIT
Generic NameCENTRAL LINE ACCESS PREP KIT
Product CodeOWL
Date Received2015-05-21
Model NumberAMS-8023CP-1
Lot Number14L23D
Device Expiration Date2016-06-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Sequence No1
Device Event Key0
ManufacturerVYGON USA
Manufacturer AddressLANDSDALE PA 19446 US 19446

Device Sequence Number: 2

Brand NameMAXPLUS CLEAR NEEDLESS CONNECTOR
Generic NameNEEDLESS CONNECTOR
Product CodeFPA
Date Received2015-05-21
Model NumberMP1000-C
Lot Number15028062
Device Sequence No2
Device Event Key0
ManufacturerCAREFUSION


Patients

Patient NumberTreatmentOutcomeDate
10 2015-05-21

© 2026 FDA.report