C SERIES CARRY BAR- SS 360755

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,08 report with the FDA on 2015-05-26 for C SERIES CARRY BAR- SS 360755 manufactured by Prism Medical.

Event Text Entries

[5916752] The patient was put in a sling and was attached to a ceiling lift. When the patient was about 1 1/2 ft. Above the bed, the carry bar cracked and broke, causing the patient to drop into bed. The carry bar fell onto the patient's abdomen. The caregiver has over 30 years of nursing experience.
Patient Sequence No: 1, Text Type: D, B5


[13389247] This incident is still under investigation.
Patient Sequence No: 1, Text Type: N, H10


[27246829] The broken plastic insert was reconstructed with glue and reconstructed part indicated deeper than normal indentation on inside face of plastic inserts, indicating use at a higher load (more than 60-70% of swl) or repeated lifting at an angle. Conclusions/recommendations: recommend ongoing evaluation of (b)(4) rev b carry bars with molded inserts to assess and further evaluate the effect of long term use on the molded inserts especially with units used at higher loads or repeatedly exposed to angular lifting during their service life.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3007802293-2015-00014
MDR Report Key4812349
Report Source01,08
Date Received2015-05-26
Date of Report2015-09-29
Date of Event2015-04-23
Date Mfgr Received2015-04-24
Device Manufacturer Date2010-07-01
Date Added to Maude2015-06-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactSTEVE KILBURN
Manufacturer Street10888 METRO COURT
Manufacturer CityMARYLAND HEIGHTS MO 63043
Manufacturer CountryUS
Manufacturer Postal63043
Manufacturer Phone3142198614
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameC SERIES CARRY BAR- SS
Generic NameCARRY BAR
Product CodeFMR
Date Received2015-05-26
Returned To Mfg2015-05-15
Model Number360755
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerPRISM MEDICAL
Manufacturer AddressMARYLAND HEIGHTS MO US


Patients

Patient NumberTreatmentOutcomeDate
10 2015-05-26

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.