SAFETY STATION PLUS TP-SS200

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2015-06-01 for SAFETY STATION PLUS TP-SS200 manufactured by Dornoch Medical.

Event Text Entries

[5834892] It was initially reported that smoke was coming from the back of the safety station plus and the unit displayed error code 270106. Additional clinical follow-up identified that the unit began smoking at the start of the cleaning cycle. The hot water master valve was getting hot and smoking while there were visible burning wires and charred plastic. The customer shut off the unit completely and the sterile processing department used an alternative method to process the waste. It was confirmed that there was no harm involved.
Patient Sequence No: 1, Text Type: D, B5


[13349328] The safety station plus was installed on (b)(4) 2008 and was last repaired on (b)(4) 2012 for a non-related issue. The device was not made available for eval or repair by zimmer surgical. The customer canceled service/repair on this device as they planned to order a new device. The customer's reported event could not be confirmed and a cause could not be determined. Please note that the device is overdue for preventative maintenance. Failure to follow the preventative maintenance schedule could lead to failures caused by normal use or wear.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1954182-2015-00001
MDR Report Key4814533
Report Source05
Date Received2015-06-01
Date of Report2015-05-18
Date of Event2015-03-03
Date Mfgr Received2015-05-18
Device Manufacturer Date2008-05-01
Date Added to Maude2015-06-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactKATHLEEN SMITH
Manufacturer Street200 NW PKWY RD
Manufacturer CityRIVERSIDE MO 64150
Manufacturer CountryUS
Manufacturer Postal64150
Manufacturer Phone3303649447
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSAFETY STATION PLUS
Generic NameSAFETY STATION PLUS
Product CodeFLH
Date Received2015-06-01
Model NumberNA
Catalog NumberTP-SS200
Lot NumberNA
ID NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerDORNOCH MEDICAL
Manufacturer Address200 NW PKWY RD RIVERSIDE MO 64150 US 64150


Patients

Patient NumberTreatmentOutcomeDate
10 2015-06-01

© 2026 FDA.report