MALYUGIN RING SYSTEM MAL-0002-1 MAL-0002

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,06 report with the FDA on 2015-06-04 for MALYUGIN RING SYSTEM MAL-0002-1 MAL-0002 manufactured by Microsurgical Technology Inc..

Event Text Entries

[19442807] After using the malyugin ring, the surgeon demonstrated to a resident how to use it properly when a piece of the ring broke off. There was no patient contact when the piece broke off and it did not cause any harm to the patient as it was not near the patient when this happened.
Patient Sequence No: 1, Text Type: D, B5


[19697731] The malyugin ring is labeled as a single use device. The surgeon was attempting to use the ring in a demonstration after the ring had already been used in surgery and had been removed from the patient and surgical field at which point the glue joint failed. The device performed according to specifications during the initial use in surgery. There was no impact to the patient.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3019924-2015-00018
MDR Report Key4819108
Report Source00,06
Date Received2015-06-04
Date of Report2015-06-03
Date of Event2015-03-03
Date Mfgr Received2015-05-11
Device Manufacturer Date2015-01-12
Date Added to Maude2015-07-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. ROBERT MAY
Manufacturer Street8415 154TH AVE NE
Manufacturer CityREDMOND WA 98052
Manufacturer CountryUS
Manufacturer Postal98052
Manufacturer Phone4255560544
Manufacturer G1MICROSURGICAL TECHNOLOGY INC
Manufacturer Street8415 154TH AVE NE
Manufacturer CityREDMOND WA 98052
Manufacturer CountryUS
Manufacturer Postal Code98052
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMALYUGIN RING SYSTEM
Generic NameIRIS CLIP RETRACTOR
Product CodeHOC
Date Received2015-06-04
Returned To Mfg2015-05-21
Model NumberMAL-0002-1
Catalog NumberMAL-0002
Lot Number066138
Device Expiration Date2018-01-01
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMICROSURGICAL TECHNOLOGY INC.
Manufacturer Address8415 154TH AVE NE REDMOND WA 98052 US 98052


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2015-06-04

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.