MOBIUS POWER *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2015-05-15 for MOBIUS POWER * manufactured by Enovate Medical.

Event Text Entries

[5839500] This is the nurse advocate charging system. Battery or battery charger sparked and caught fire. Battery was destroyed and unable to read serial number. Serial number of charger noted in report. Two charging stations damaged as a result. Like models of batteries lot #'s: 311801313021259 in adjacent charging station; 311801313050208 next to one catching fire. Other serial numbers not readable, model # mobius is 8116270. Size of battery 10. 95 volts, 26. 5 amp hours, 290. 18 watt hours. ======================manufacturer response for battery and battery charger, stinger medical mobius power (per site reporter). ======================field tech from ennovate medical on site. Field tech updated firmware of charging stations throughout hospital. Five charging stations total that were updated.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number4823868
MDR Report Key4823868
Date Received2015-05-15
Date of Report2015-04-28
Date of Event2015-04-19
Report Date2015-04-28
Date Reported to FDA2015-05-15
Date Reported to Mfgr2015-06-08
Date Added to Maude2015-06-08
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag0
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameMOBIUS POWER
Generic NameBATTERY CHARGER
Product CodeFCO
Date Received2015-05-15
Model Number*
Catalog Number*
Lot Number*
ID Number*
Device AvailabilityY
Device Age18 MO
Device Sequence No1
Device Event Key0
ManufacturerENOVATE MEDICAL
Manufacturer Address1152 PARK AVE. MURFREESBORO TN 37129 US 37129


Patients

Patient NumberTreatmentOutcomeDate
10 2015-05-15

© 2026 FDA.report