TC MAYO HEGAR NDL HOLDERHVYSERR205MM BM067R

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07,company representative report with the FDA on 2015-06-10 for TC MAYO HEGAR NDL HOLDERHVYSERR205MM BM067R manufactured by Aesculap Ag & Co. Kg.

Event Text Entries

[5932226] Tip of tc insert broke during c-section operation. No pt injury. Surgery delay 40-45 mins. Broken tip was retrieved with direct visualization, x-ray used to confirm no fragments remained. The broken piece of the tip will not be returned for evaluation.
Patient Sequence No: 1, Text Type: D, B5


[13353300] Manufacturing site evaluation: investigation on-going.
Patient Sequence No: 1, Text Type: N, H10


[35325928] (b)(4). Manufacturing site evaluation: the tip of the carbide metal insert of the received device is broken off. The instrument was analyzed with the digital microscope (b)(4) by keyence. Hardness was tested with hardness tester vickers hv 5, ident no. 30077. Carbide metal - target:1150 +-50 hv , actual: 1180 hv. Basic material - target: 420 +110 hv , actual: 505 hv. The hardness is according to the specifications. It is assumed that the user tried to grasp a needle with the tips of the needle holder as the needle slipped and caused the breakage; however this is not confirmed. The damage is likely user related. Corrective / preventive action(s) is not required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2916714-2015-00480
MDR Report Key4845224
Report Source07,COMPANY REPRESENTATIVE
Date Received2015-06-10
Date of Report2015-12-30
Date Mfgr Received2015-05-12
Date Added to Maude2015-06-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactMS. NICOLE BROYLES
Manufacturer Street615 LAMBERT POINTE DRIVE
Manufacturer CityHAZELWOOD MO 63042
Manufacturer CountryUS
Manufacturer Postal63042
Manufacturer Phone3145515988
Manufacturer G1AESCULAP AG & CO KG
Manufacturer StreetP.O. BOX 40
Manufacturer CityTUTTLINGEN 78501
Manufacturer CountryGM
Manufacturer Postal Code78501
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTC MAYO HEGAR NDL HOLDERHVYSERR205MM
Generic NameNEEDLE HOLDER
Product CodeHXK
Date Received2015-06-10
Model NumberBM067R
Catalog NumberBM067R
OperatorHEALTH PROFESSIONAL
Device Availability*
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerAESCULAP AG & CO. KG
Manufacturer AddressTUTTLINGEN, DE 78532 78532


Patients

Patient NumberTreatmentOutcomeDate
10 2015-06-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.