[5872977] The customer received questionable aspartate aminotransferase (ast) results for one patient from a cobas c501. (b)(6) 2015: 311 u/l 02/21/2015: 551 u/l 10/09/2013: 291 u/l it was uncertain if these results were generated from the same patient sample. With the sample that generated the result of 311 u/l, the customer performed precipitation with peg5000: sample with peg5000 25% in pbs (1)- <10 u/l sample with peg5000 25% in pbs (2)- <10 u/l %ppa 96. 8% (ppa ref 18-53 %). The results were reported outside the laboratory. The patient was not adversely affected. The customer suspected the presence of a macroenzyme, one complexed with an immunoglobulin and ast.
Patient Sequence No: 1, Text Type: D, B5


[13481746] This event occurred in (b)(6).
Patient Sequence No: 1, Text Type: N, H10


[22562626] A specific root cause could not be identified. Based on the provided data, it was determined the test measured the correct amount of ast in the patient sample which was elevated due to possible macro enzymes.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1823260-2015-03640
MDR Report Key4846629
Report Source01,05,06
Date Received2015-06-16
Date of Report2015-08-04
Date of Event2015-04-06
Date Mfgr Received2015-05-28
Date Added to Maude2015-06-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactNA MICHAEL LESLIE
Manufacturer Street9115 HAGUE ROAD NA
Manufacturer CityINDIANAPOLIS IN 46250
Manufacturer CountryUS
Manufacturer Postal46250
Manufacturer Phone3175214343
Manufacturer G1ROCHE DIAGNOSTICS GMBH
Manufacturer StreetSANDHOFERSTRASSE 116 NA
Manufacturer CityMANNHEIM (BADEN-WURTTEMBERG) 68305
Manufacturer CountryGM
Manufacturer Postal Code68305
Single Use3
Previous Use Code3
Removal Correction NumberNA
Event Type3
Type of Report3

Device Details

Brand NameASPARTATE AMINOTRANSFERASE ACC. TO IFCC WITHOUT PYRIDOXAL PHOSPHATE ACTIVATION
Generic NameNADH OXIDATION/NAD REDUCTION, AST/SGOT
Product CodeCIT
Date Received2015-06-16
Model NumberNA
Catalog Number20764949322
Lot Number61168
ID NumberNA
Device Expiration Date2016-05-30
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerROCHE DIAGNOSTICS
Manufacturer Address9115 HAGUE ROAD NA INDIANAPOLIS IN 46250 US 46250


Patients

Patient NumberTreatmentOutcomeDate
10 2015-06-16

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.